Journal of the Brazilian Chemical Society | |
Laboratory validation of a passive sampler for SO2 atmospheric monitoring | |
Cruz, Lícia P. S.1  Universidade Federal da Bahia, Salvador, Brazil1  Tavares, Tania M.1  Campos, Vânia P.1  Novaes, Jussival A. P.1  | |
关键词: passive sampling; atmospheric sulfur dioxide; air monitoring; | |
DOI : 10.1590/S0103-50532005000100008 | |
学科分类:化学(综合) | |
来源: SciELO | |
【 摘 要 】
The performance of a passive sampler was tested under exposure in Teflon chamber. The SO2 trapped in impregnated filter with 1.0 x 10-2 mol L-1 Na2CO3 solution was extracted in ultrasound bath with H2O2 solution for 15 min and determined as SO42- by ion chromatography. The sampler presented good performance except for air in total stillness condition (0 m s-1), for very short exposure periods (0.5 h), for atmosphere with O3 > 150 mg m-3 and for high concentrations of reduced sulfur compounds. The relative humidity of the air (20 - 80%) and the exposure to zero concentration did not affect the sampler performance. It showed stability for one month at 4 ºC or for one week at about 25 ºC. Its precision varied within 3.5 and 23% depending of the SO2 concentration and the accuracy varied from 91 to 100%.