Arquivos Brasileiros de Oftalmologia
Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging
Fernando Faria Correia2  Isaac Ramos2  Cynthia J. Roberts1  Andreas Steinmueller1  Matthias Krug1  Renato Ambrósio Jr.2 
[1] ,Instituto de Olhos Renato AmbrósioRio de Janeiro RJ ,Brazil
关键词: Cornea;    Corneal topography;    Biomechanics;    Tonometry;    ocular;    Pressure;    Córnea;    Topografia da córnea;    Biomecância;    Tonometria ocular;    Pressão;   
DOI  :  10.1590/S0004-27492013000500005
来源: SciELO
PDF
【 摘 要 】

PURPOSE: To study the deformation response of three distinct contact lenses with known structures, which served as corneal models, under different chamber pressures using ultra-high-speed (UHS) Scheimpflug imaging. METHODS: Three hydrophilic contact lenses were mounted on a sealed water chamber with precisely adjustable pressure: TAN-G5X (41% hydroxyethylmethacrylate/glycolmethacrylate, 550 µm thick), TAN-40 (62% hydroxyethylmethacrylate, 525 µm thick) and TAN-58 (42% methylmethacrylate, 258 µm thick). Each model was tested five times under different pressures (5, 15, 25, 35 and 45 mmHg), using ultra-high-speed Scheimpflug imaging during non-contact tonometry. 140 Scheimpflug images were taken with the UHS camera in each measurement. The deformation amplitude during non-contact tonometry was determined as the highest displacement of the apex at the highest concavity (HC) moment. RESULTS: At each pressure level, the deformation amplitude was statistically different for each lens tested (p<0.001, ANOVA). Each lens had different deformation amplitudes under different pressure levels (p<0.001; Bonferroni post-hoc test). The thicker lens with less polymer (TAN-G5X) had a higher deformation (less stiff behavior) than the one that was thinner but with more polymer (TAN-40), when measured at the same internal pressure. The thinnest lens with less polymers (TAN-58) had a lower deformation amplitude (stiffer behavior) at higher pressures than the thicker ones with more polymer (TAN-40 and TAN-G5X) at lower pressures. CONCLUSIONS: UHS Scheimpflug imaging allowed for biomechanical assessment through deformation characterization of corneal models. Biomechanical behavior was more influenced by material composition than by thickness. Chamber pressure had a significant impact on deformation response of each lens.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130003573ZK.pdf 338KB PDF download
  文献评价指标  
  下载次数:9次 浏览次数:12次