期刊论文详细信息
Sensors
Study of Sodium Ion Selective Electrodes and Differential Structures with Anodized Indium Tin Oxide
Jyh-Ling Lin1 
关键词: anodized indium tin oxide;    anodic oxidation;    sodium ion;    EGFET;   
DOI  :  10.3390/s100301798
来源: mdpi
PDF
【 摘 要 】

The objective of this work is the study and characterization of anodized indium tin oxide (anodized-ITO) as a sodium ion selective electrode and differential structures including a sodium-selective-membrane/anodized-ITO as sensor 1, an anodized-ITO membrane as the contrast sensor 2, and an ITO as the reference electrode. Anodized-ITO was fabricated by anodic oxidation at room temperature, a low cost and simple manufacture process that makes it easy to control the variation in film resistance. The anodized-ITO based on EGFET structure has good linear pH sensitivity, approximately 54.44 mV/pH from pH 2 to pH 12. The proposed sodium electrodes prepared by PVC-COOH, DOS embedding colloid, and complex Na-TFBD and ionophore B12C4, show good sensitivity at 52.48 mV/decade for 10−4 M to 1 M, and 29.96 mV/decade for 10−7 M to 10−4 M. The sodium sensitivity of the differential sodium-sensing device is 58.65 mV/decade between 10−4 M and 1 M, with a corresponding linearity of 0.998; and 19.17 mV/decade between 10−5 M and 10−4 M.

【 授权许可】

CC BY   
© 2010 by the authors; licensee Molecular Diversity Preservation International, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190054840ZK.pdf 230KB PDF download
  文献评价指标  
  下载次数:3次 浏览次数:4次