期刊论文详细信息
International Journal of Environmental Research and Public Health
Assessment of Electromagnetic Interference with Active Cardiovascular Implantable Electronic Devices (CIEDs) Caused by the Qi A13 Design Wireless Charging Board
Tobias Seckler1  Kai Jagielski2  Dominik Stunder2 
[1] The Research Center for Bioelectromagnetic Interaction (FEMU), Institute and Out-patient Clinic of Occupational Medicine, RWTH Aachen University, Pauwelsstr 30, 52074 Aachen, Germany;
关键词: intermediate frequency magnetic fields;    pacemaker;    defibrillator;    electromagnetic interference;    wireless charging;    wireless power transfer;    Qi A13 design;    performance limits;    EMF risk assessment;   
DOI  :  10.3390/ijerph120605886
来源: mdpi
PDF
【 摘 要 】

Electromagnetic interference is a concern for people wearing cardiovascular implantable electronic devices (CIEDs). The aim of this study was to assess the electromagnetic compatibility between CIEDs and the magnetic field of a common wireless charging technology. To do so the voltage induced in CIEDs by Qi A13 design magnetic fields were measured and compared with the performance limits set by ISO 14117. In order to carry this out a measuring circuit was developed which can be connected with unipolar or bipolar pacemaker leads. The measuring system was positioned at the four most common implantation sites in a torso phantom filled with physiological saline solution. The phantom was exposed by using Helmholtz coils from 5 µT to 27 µT with 111 kHz sine-bursts or by using a Qi A13 design wireless charging board (Qi-A13-Board) in two operating modes “power transfer” and “pinging”. With the Helmholtz coils the lowest magnetic flux density at which the performance limit was exceeded is 11 µT. With the Qi-A13-Board in power transfer mode 10.8% and in pinging mode 45.7% (2.2% at 10 cm distance) of the performance limit were reached at maximum. In neither of the scrutinized cases, did the voltage induced by the Qi-A13-Board exceed the performance limits.

【 授权许可】

CC BY   
© 2015 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190012020ZK.pdf 1998KB PDF download
  文献评价指标  
  下载次数:14次 浏览次数:59次