Sensors | |
Design and Evaluation of Potentiometric Principles for Bladder Volume Monitoring: A Preliminary Study | |
Shih-Ching Chen1  Tsung-Hsun Hsieh2  Wen-Jia Fan1  Chien-Hung Lai1  Chun-Lung Chen1  Wei-Feng Wei3  Chih-Wei Peng1  | |
[1] Department of Physical Medicine and Rehabilitation, School of Medicine, College of Medicine, Taipei Medical University, No. 250, Wuxing Street, Taipei 11031, Taiwan; E-Mails:;Graduate Institute of Neural Regenerative Medicine, Taipei Medical University, No. 250, Wuxing Street, Taipei 11031, Taiwan; E-Mail:;Graduate Institute of Biomedical Electronics and Bioinformatics, National Taiwan University, No.1, Roosevelt Road, Sec.4, Taipei 10617, Taiwan; E-Mail: | |
关键词: potentiometer; real time; pig bladder; implantable; sensor; | |
DOI : 10.3390/s150612802 | |
来源: mdpi | |
【 摘 要 】
Recent advances in microelectronics and wireless transmission technology have led to the development of various implantable sensors for real-time monitoring of bladder conditions. Although various sensing approaches for monitoring bladder conditions were reported, most such sensors have remained at the laboratory stage due to the existence of vital drawbacks. In the present study, we explored a new concept for monitoring the bladder capacity on the basis of potentiometric principles. A prototype of a potentiometer module was designed and fabricated and integrated with a commercial wireless transmission module and power unit. A series of
【 授权许可】
CC BY
© 2015 by the authors; licensee MDPI, Basel, Switzerland.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202003190011634ZK.pdf | 3495KB | download |