Sensors
A Multi-Channel Opto-Electronic Sensor to Accurately Monitor Heart Rate against Motion Artefact during Exercise
Abdullah Alzahrani1  Sijung Hu1  Vicente Azorin-Peris1  Laura Barrett5  Dale Esliger5  Matthew Hayes2  Shafique Akbare3  Jérôme Achart3  Sylvain Kuoch3  Steffen Leonhardt4 
[1] Wolfson School of Mechanical, Manufacturing and Electrical Engineering, Loughborough University, Ashby Road, Loughborough, Leicestershire LE11 3TU, UK; E-Mails:;Cambridge Consultants Limited, Science Park, Milton Road, Cambridge CB4 0DW, UK; E-Mail:;Université Paris-Sud 11, Polytech’Paris-Sud–Service des stages, Bât 620, Orsaycedex 91405, France; E-Mails:Wolfson School of Mechanical, Manufacturing and Electrical Engineering, Loughborough University, Ashby Road, Loughborough, Leicestershire LE11 3TU, UK;;National Centre for Sport and Exercise Medicine, School of Sport, Exercise and Health Sciences, Loughborough University, Ashby Road, Loughborough, Leicestershire LE11 3TU, UK; E-Mails:
关键词: multi-channel opto-electronics;    accelerometer;    motion artefact;    sport physiological monitoring;    heart rate;    pulse wave;    vital signs;    wearables;   
DOI  :  10.3390/s151025681
来源: mdpi
PDF
【 摘 要 】

This study presents the use of a multi-channel opto-electronic sensor (OEPS) to effectively monitor critical physiological parameters whilst preventing motion artefact as increasingly demanded by personal healthcare. The aim of this work was to study how to capture the heart rate (HR) efficiently through a well-constructed OEPS and a 3-axis accelerometer with wireless communication. A protocol was designed to incorporate sitting, standing, walking, running and cycling. The datasets collected from these activities were processed to elaborate sport physiological effects. t-test, Bland-Altman Agreement (BAA), and correlation to evaluate the performance of the OEPS were used against Polar and Mio-Alpha HR monitors. No differences in the HR were found between OEPS, and either Polar or Mio-Alpha (both p > 0.05); a strong correlation was found between Polar and OEPS (r: 0.96, p < 0.001); the bias of BAA 0.85 bpm, the standard deviation (SD) 9.20 bpm, and the limits of agreement (LOA) from −17.18 bpm to +18.88 bpm. For the Mio-Alpha and OEPS, a strong correlation was found (r: 0.96, p < 0.001); the bias of BAA 1.63 bpm, SD 8.62 bpm, LOA from −15.27 bpm to +18.58 bpm. These results demonstrate the OEPS to be capable of carrying out real time and remote monitoring of heart rate.

【 授权许可】

CC BY   
© 2015 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190005118ZK.pdf 2703KB PDF download
  文献评价指标  
  下载次数:41次 浏览次数:26次