期刊论文详细信息
Sensors
A Spectrometric Method for Hydrogen Peroxide Concentration Measurement with a Reusable and Cost-Efficient Sensor
Cheng-Chih Hsu1  Yuan-Rong Lo1  Yu-Chian Lin1  Yi-Cen Shi1  Pang-Lung Li1 
关键词: optical sensor;    spectrometry;    hydrogen peroxide;   
DOI  :  10.3390/s151025716
来源: mdpi
PDF
【 摘 要 】

In this study we developed a low cost sensor for measuring the concentration of hydrogen peroxide (H2O2) in liquids utilizing a spectrometric method. The sensor was tested using various concentrations of a peroxidase enzyme immobilized on a glass substrate. H2O2 can be catalyzed by peroxidase and converted into water and oxygen. The reagent 4-amino-phenazone takes up oxygen together with phenol to form a colored product that has absorption peaks at 510 nm and 450 nm. The transmission intensity is strongly related to the hydrogen peroxide concentration, so can be used for quantitative analysis. The measurement range for hydrogen peroxide is from 5 × 105% to 1 × 10−3% (0.5 ppm to 10 ppm) and the results show high linearity. This device can achieve a sensitivity and resolution of 41,400 (photon count/%) and 3.49 × 10−5% (0.35 ppm), respectively. The response time of the sensor is less than 3 min and the sensor can be reused for 10 applications with similar performance.

【 授权许可】

CC BY   
© 2015 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190005116ZK.pdf 1269KB PDF download
  文献评价指标  
  下载次数:7次 浏览次数:4次