期刊论文详细信息
Metrology and Measurement Systems
Laser Reflectance Interferometry System with a 405 Nm Laser Diode for in Situ Measurements of CVD Diamond Thickness
Robert Bogdanowicz1  Maciej Kraszewski1 
关键词: interferometry;    thin films;    CVD;    diamond;    in situ measurements;   
DOI  :  10.2478/mms-2013-0046
来源: Versita
PDF
【 摘 要 】

In situ monitoring of the thickness of thin diamond films during technological processes is important because it allows better control of deposition time and deeper understanding of deposition kinetics. One of the widely used techniques is laser reflectance interferometry (LRI) which enables non-contact measurement during CVD deposition. The authors have built a novel LRI system with a 405 nm laser diode which achieves better resolution compared to the systems based on He-Ne lasers, as reported so far. The system was used for in situ monitoring of thin, microcrystalline diamond films deposited on silicon substrate in PA-CVD processes. The thickness of each film was measured by stylus profilometry and spectral reflectance analysis as a reference. The system setup and interferometric signal processing are also presented for evaluating the system parameters, i.e. measurement uncertainty, resolution and the range of measurable film thickness.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912080706970ZK.pdf 482KB PDF download
  文献评价指标  
  下载次数:15次 浏览次数:49次