期刊论文详细信息
Metrology and Measurement Systems
Ageing of Cadmium Telluride Radiation Detectors and its Diagnostics with Low Frequency Noise
Lubomir Grmela1  Alexey Andreev1  Josef Sikula1  Ondrej Sik1 
关键词: noise spectroscopy;    CdTe radiation detectors;    ageing process;   
DOI  :  10.2478/mms-2013-0033
来源: Versita
PDF
【 摘 要 】

Samples of CdTe single crystals which are used as radiation detectors were periodically measured during a long time interval with different values of an applied voltage. The samples were also periodically exposed during long time periods to high temperatures of 390 K and to rapid changes of temperature from 300 K to 390 K. After 1.5 years of measurements we observed ageing of the samples which resulted in deterioration of their transport characteristics. The resistance of the samples increased significantly and current-voltage characteristics were unstable in time. Noise spectroscopy showed that low frequency noise can be used for detection of CdTe sample ageing as its spectral density increases significantly comparing to the 1/f noise of a high quality sample

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912080706957ZK.pdf 460KB PDF download
  文献评价指标  
  下载次数:5次 浏览次数:21次