期刊论文详细信息
JOURNAL OF CHEMICAL ENGINEERING OF JAPAN
Micro-Flow System Comprised of a Fused-Silica Capillary and Chemiluminescence Detection that Works under Laminar Flow Conditions
Riichiro Nakajima1  Kazuhiko Tsukagoshi1  Shingo Ishida1 
[1] Department of Chemical Engineering and Materials Science, Faculty of Engineering, Doshisha University
关键词: Micro-Flow Analysis;    Laminar Flow;    Fused-Silica Capillary;    Chemiluminescence;   
DOI  :  10.1252/jcej.07WE216
来源: Maruzen Company Ltd
PDF
【 摘 要 】

References(13)Cited-By(4)A micro-flow system comprised of a fused-silica capillary and chemiluminescence detection was developed, in which a wide-bore hydrodynamic chromatography mode was first introduced into the system. The micro-flow system was run with a gravity method or pump flow method for feeding a carrier solution into the capillary. Oxalate–hydrogen peroxide chemiluminescence reaction was adopted as a detection technique. Fluorescent compound (fluorescein isothiocyanate; FITC) and FluoSphere fluorescent polymer particles with particle diameters of 0.02 μm and 1.0 μm were used as model samples in the present system. FITC and FluoSpheres (0.02 μm) with diffusion indexes greater than ca. 1 were delivered into the capillary with average linear velocities under laminar flow conditions, which led to a Gaussian peak on the chemiluminescence profile. On the other hand, FluoSpheres (1.0 μm) with a diffusion index of ca. 0.1 were delivered with the approximately maximum linear velocity and formed a non-Gaussian peak. The mixtures, FITC and FluoSpheres (1.0 μm) as well as FluoSpheres (0.02 μm) and FluoSpheres (1.0 μm), reflected the individual diffusion indexes, indicating two main peaks on the chemiluminescence profiles. An unexpected small Gaussian peak observed with the FluoSpheres (1.0 μm) sample was also observed.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912080696180ZK.pdf 19KB PDF download
  文献评价指标  
  下载次数:14次 浏览次数:14次