| Química Nova | |
| Use of the HRICP-MS technique for the evaluation of boron isotopes in eucalitpus plants | |
| Mattiello, Edson Marcio1  Instituto de Pesquisas Energética e Nucleares, São Paulo, Brasil1  Universidade Federal de Viçosa, Viçosa, Brasil1  Ruiz, Hugo Alberto1  Sarkis, Jorge Eduardo de Souza1  Silva, Ivo Ribeiro da1  | |
| 关键词: isotope ratio; micronutrients; translocation.; | |
| DOI : 10.1590/S0100-40422011000300025 | |
| 学科分类:化学(综合) | |
| 来源: Sociedade Brasileira de Quimica | |
PDF
|
|
【 摘 要 】
The 10B isotope tracer technique is essential to study the B mobility in plants. Factors that can influence the quality of measured B isotope ratios were optimized experimentally using High Resolution Inductively Coupled Plasma Mass Spectrometry (HRICP-MS). An isotopically certified standard (NIST SRM-951) was used. The best combination was obtained using a resolution of 400, a RF power of 1250 W, followed by 15 measurements over a 10-s integration period each (15*10). Utilizing this approach it was possible to obtain a precision of 0.3 % in standard material and 2z % in the experimental samples. The results show the importance of establishing optimized work conditions before carrying out the analytical series.
【 授权许可】
Unknown