| Uso da técnica HRICP-MS na avaliação dos isótopos de boro em eucalipto | |
| Edson Marcio Mattiello2  Hugo Alberto Ruiz2  Ivo Ribeiro Da Silva2  Jorge Eduardo De Souza Sarkis1  | |
| [1] ,Universidade Federal de Viçosa Departamento de Solos Viçosa MG ,Brasil | |
| 关键词: isotope ratio; micronutrients; translocation; | |
| DOI : 10.1590/S0100-40422011000300025 | |
| 来源: SciELO | |
PDF
|
|
【 摘 要 】
The 10B isotope tracer technique is essential to study the B mobility in plants. Factors that can influence the quality of measured B isotope ratios were optimized experimentally using High Resolution Inductively Coupled Plasma Mass Spectrometry (HRICP-MS). An isotopically certified standard (NIST SRM-951) was used. The best combination was obtained using a resolution of 400, a RF power of 1250 W, followed by 15 measurements over a 10-s integration period each (15*10). Utilizing this approach it was possible to obtain a precision of 0.3 % in standard material and 2z % in the experimental samples. The results show the importance of establishing optimized work conditions before carrying out the analytical series.