Química Nova | |
Determination of sildenafil citrate and tadalafil by ultra performance liquid chromatography with diode-array detection (UPLC-DAD) | |
Linden, Rafael1  Centro Universitário Feevale, Novo Hamburgo, Brasil1  Antunes, Marina Venzon1  Ortiz, Rafael Scorsatto1  Superintendência de Polícia Federal no Estado do Rio Grande do Sul, Porto Alegre, Brasil1  | |
关键词: sildenafil; tadalafil; ultra performance liquid chromatography.; | |
DOI : 10.1590/S0100-40422010000200028 | |
学科分类:化学(综合) | |
来源: Sociedade Brasileira de Quimica | |
【 摘 要 】
A simple ultra-performance liquid chromatographic method for the simultaneous determination of sildenafil citrate and tadalafil was developed and validated. Sample preparation was dissolution in methanol, followed by centrifugation and dilution (1:10) with methanol. Analysis was performed in an Acquity® UPLC system with Acquity® BEH C18 column (2.1 x 50 mm, with 1.7 μm particles). The elution was isocratic with phosphate buffer pH 2.3 and acetonitrile (65:35, v/v) at a flow rate of 0.7 mL/min. The method presented adequate specificity, linearity, precision and accuracy and allowed the determination of the drugs in seized forensic samples.
【 授权许可】
Unknown
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201912050594879ZK.pdf | 229KB | download |