期刊论文详细信息
Química Nova
Evaluation of analytical methods for BTEX analysis in water using extraction by headspace (HS) and solid phase microextraction (SPME)
Heleno, Fernanda F.1  Universidade Federal de Ouro Preto, Ouro Preto, Brasil1  Lima, Aniel C.1  Coutrim, Mauricio X.1  Afonso, Robson J. C. F.1 
关键词: SPME;    headspace;    BTEX.;   
DOI  :  10.1590/S0100-40422010000200019
学科分类:化学(综合)
来源: Sociedade Brasileira de Quimica
PDF
【 摘 要 】

Three analytical methods for the determination of BTEX in water were optimized and validated. With the best method the analytes were extracted of 10 mL of sample with 2.50 g of NaCl in headspace vial of 20 mL by HS and SPME to 40 ºC for 30 min for adsorption and to 250 ºC for 4 min for desorption and were analyzed by GC-MS. The recovery was between 97.9% and 104.3%, and the limit of detection was 2.4 ng L-1 for o-xylene. This method was using to analyze BTEX in water supply and surface water in Ouro Preto city. No sample had concentrations of BTEX above the legislation.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912050594870ZK.pdf 832KB PDF download
  文献评价指标  
  下载次数:6次 浏览次数:5次