Química Nova | |
Photomultiplier nonlinear response in time-domain laser-induced luminescence spectroscopy | |
Schip, Leandro José Bossy1  Buzelatto, Bruno Phelippe1  Dias Jr., Lauro Camargo1  Universidade Federal do Paraná, Curitiba, Brasil1  Batista, Fábio Roberto1  Novo, João Batista Marques1  Cunha, Carlos Jorge da1  | |
关键词: photomultiplier; time-resolved; laser-induced luminescence.; | |
DOI : 10.1590/S0100-40422007000100034 | |
学科分类:化学(综合) | |
来源: Sociedade Brasileira de Quimica | |
【 摘 要 】
A new procedure to find the limiting range of the photomultiplier linear response of a low-cost, digital oscilloscope-based time-resolved laser-induced luminescence spectrometer (TRLS), is presented. A systematic investigation on the instrument response function with different signal input terminations, and the relationship between the luminescence intensity reaching the photomultiplier and the measured decay time are described. These investigations establish that setting the maximum intensity of the luminescence signal below 0.3V guarantees, for signal input terminations equal or higher than 99.7 ohm, a linear photomultiplier response.