【 摘 要 】
A new procedureto find the limiting range of the photomultiplier linear response of a low-cost,digital oscilloscope-based time-resolved laser-induced luminescence spectrometer(TRLS), is presented. A systematic investigation on the instrument responsefunction with different signal input terminations, and the relationship betweenthe luminescence intensity reaching the photomultiplier and the measured decaytime are described. These investigations establish that setting the maximumintensity of the luminescence signal below 0.3V guarantees, for signal inputterminations equal or higher than 99.7 ohm, a linear photomultiplier response.