期刊论文详细信息
Journal of the Brazilian Chemical Society
Validation method to determine metals in atmospheric particulate matter by inductively coupled plasma optical emission spectrometry
Gioda, Adriana1  Ventura, Luciana M. B.1  Amaral, Beatriz S.1  Pontifical Catholic University of Rio de Janeiro (PUC-Rio), Rio de Janeiro, Brazil1  Godoy, José M.1  Wanderley, Kristine B.1 
关键词: validation method;    PM2.5;    ICP-OES;    atmospheric chemistry;   
DOI  :  10.5935/0103-5053.20140142
学科分类:化学(综合)
来源: SciELO
PDF
【 摘 要 】

Atmospheric particulate matter (PM) is a pollutant composed by various metals, that when accumulated in the respiratory system may cause serious health problems. Methods IO-3.1 (metal extraction in PM) and IO-3.4 [metal determination by inductively coupled plasma optical emission spectrometry (ICP-OES)] are the recommended by the United State Environmental Protection Agency. With the intent to evaluate the performance of the method developed in our laboratory for the extraction of metal in PM with HNO3 p.a. bidistilled and determination by ICP-OES of Al, Ca, Cd, Cu, Cr, Fe, K, Mg, Mn, Na, Ni, Pb, Ti, V and Zn, validation was executed according to the criteria established by the INMETRO determining the parameters: selectivity, linearity, precision, accuracy, robustness, limits of detection and quantification, as well as the comparison of the efficiency and precision with IO-3.1 method. The results show that our method meets all validation criteria established by the INMETRO; furthermore, it shows to be equivalent to IO-3.1 method.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912050582267ZK.pdf 503KB PDF download
  文献评价指标  
  下载次数:8次 浏览次数:14次