期刊论文详细信息
Journal of the Brazilian Chemical Society
Quantification of autoantibodies to annexin V in plasma by an "in house" sandwich ELISA
Arriaga, Sandra M.1  Daniele, Stella M.1  Pelusa, Héctor F.1  Bearzotti, Mariela1  Ghersevich, Sergio A.1  Valdés, María de los Ángeles1  Bouvet, Beatriz R.1  Almará, Adriana M.1  Paparella, Cecilia V.1  Svetaz, María J.1  Universidad Nacional de Rosario, Rosario, Argentina1 
关键词: ELISA;    annexin V;    antibodies;    anti-phospholipid syndrome;    foetal losses;   
DOI  :  10.1590/S0103-50532012000300011
学科分类:化学(综合)
来源: SciELO
PDF
【 摘 要 】

An "in house" sandwich ELISA for the quantification of plasma anti-annexin V antibodies was developed and validated. The validation parameters studied were: (i) precision, expressed either as the intra- or the inter-assay coefficient of variation (CV), (ii) exactitude, expressed as the percentage deviation between the obtained value and the real value, (iii) limit of detection (LOD), evaluated from the reagents blank and (iv) robustness, obtained by deliberately introducing slight variations in different parameters. Also, a comparison between the "in house" technique and a commercial method was performed. The research revealed that both CV were < 20%, exactitude was within the 100 ± 20% range, limit of detection was below 1 U mL-1 and that slight variations in the technique did not produce any significant variations in the results. Comparison with the commercial method showed an acceptable correlation. It was concluded that the method developed here satisfactorily accomplishes the parameters of standardization and validation for an immunoassay.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912050581572ZK.pdf 498KB PDF download
  文献评价指标  
  下载次数:8次 浏览次数:15次