期刊论文详细信息
Journal of the Brazilian Chemical Society
Determination and validation of nimesulide in pharmaceutical formulation by near infrared spectroscopy
Instituto Vita Nova, Hortolândia, Brazil1  Rocha, Werickson F. C.1  Martins, José Antônio1  University of Campinas, Campinas, Brazil1  Poppi, Ronei Jesus1  Rosa, André Luis1 
关键词: near-infrared spectroscopy;    nimesulide;    figures of merit;    net analyte signal;   
DOI  :  10.1590/S0103-50532010001000019
学科分类:化学(综合)
来源: SciELO
PDF
【 摘 要 】

The determination of nimesulide in pharmaceutical formulation in the range of 10.38 to 39.47% m/m using near infrared spectroscopy and multivariate calibration based on partial least squares (PLS) and net analyte signal (NAS) was studied and validated by establishing its figures of merit. Suitable results were obtained, with limit of detection of 0.61%, limit quantification of 2.03%, accuracy estimated as root mean square error of prediction (RMSEP) of 1.05%, mean selectivity of 0.006, sensitivity of 0.004%, inverse analytical sensitivity of (0.20%)-1 and signal-to-noise ratio of 181. In addition, it was obtained relative errors lower than 10% that are acceptable values for quality control of this drug. These results indicate that the proposed method can be used in pharmaceutical industries as an alternative to standard analytical procedures.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912050580854ZK.pdf 340KB PDF download
  文献评价指标  
  下载次数:6次 浏览次数:9次