期刊论文详细信息
Pramana
A perturbed angular correlation spectrometer for material science studies
C C Dey11 
[1] Nuclear and Atomic Physics Division, Saha Institute of Nuclear Physics, 1/AF, Bidhannagar, Kolkata 700 064, India$$
关键词: Time differential perturbed angular correlation;    BaF2-BaF2;    HfO2;    HfF4.3H2O.;   
DOI  :  
学科分类:物理(综合)
来源: Indian Academy of Sciences
PDF
【 摘 要 】

A four-detector perturbed angular correlation (PAC) spectrometer has been developed with ultra-fast BaF2 detectors to acquire four coincidence spectra simultaneously, two at 180° and two at 90°. This spectrometer has double efficiency compared to that of a three-detector set-up. Higher efficiency is desirable for PAC studies in solid state physics where large number of coincidences are required to obtain the PAC spectra with good statistics and is particularly useful when the half-lives of the parent probe nuclei used for PAC measurements are ∼ 2-3 days or less as in 111In (2.8 d), 99Mo (2.7 d) and 140La (1.7 d). The performance of the spectrometer has been tested for the HfO2 monoclinic crystal in the temperature range from 77 to 873 K and for the HfF4.3H2O crystal at room temperature. The polycrystalline HfO2 has been synthesized from Hf metal by heating in air. The hydrated hafnium fluoride has been crystallized by dissolving Hf metal in 40% HF and drying slowly at room temperature.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912040497571ZK.pdf 481KB PDF download
  文献评价指标  
  下载次数:12次 浏览次数:22次