Experimental study of the dependence of beam current on injection magnetic field in 6.4 GHz ECR ion source
D K Chakraborty2  R K Bhandari2  P K Ray1  P R Sarma2  G S Taki12 
[1] Bengal Engineering and Science University, Shibpur, Howrah 711 103, India$$;Department of Atomic Energy, Variable Energy Cyclotron Centre, Block-AF, Bidhan Nagar, Kolkata 700 064, India$$
关键词: Electron cyclotron resonance ion source;    mirror ratio;    field scaling;    ion confinement time;    ion current.;   
DOI  :  
学科分类:物理(综合)
来源: Indian Academy of Sciences
PDF
【 摘 要 】

The ion current from an electron cyclotron resonance (ECR) heavy ion source depends on the confining axial and radial magnetic fields. Some efforts were made by earlier workers to investigate magnetic field scaling on the performance of the ECR source. In order to study the dependence of the ion current on the injection magnetic field in the 6.4 GHz ECR source, we have measured the current by varying the peak injection field and have inferred that the variation of the current is exponential up to our maximum design injection field of 7.5 kG. An attempt has been made to understand this exponential nature on the basis of ion confinement time.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912040497165ZK.pdf 976KB PDF download
  文献评价指标  
  下载次数:18次 浏览次数:20次