期刊论文详细信息
Pramana
The automatic liquid nitrogen filling system for GDA detectors
Rakesh Kumar1  A J Malyadri1  R K Bhowmik1  B P Ajith Kumar1  Rajesh Kumar1  S K Saini Kusum Rani1  Ruby Shanti1  S Muralithar1 
[1] Nuclear Science Centre, Aruna Asaf Ali Marg, Post Box No 10502, New Delhi 110 067, India$$
关键词: Automation;    liquid nitrogen;    high purity germanium;    electro-pneumatic.;   
DOI  :  
学科分类:物理(综合)
来源: Indian Academy of Sciences
PDF
【 摘 要 】

An indigenously developed automatic liquid nitrogen (LN2) filling system has been installed in gamma detector array (GDA) facility at Nuclear Science Centre. Electro-pneumatic valves are used for filling the liquid nitrogen into the high purity germanium detector cryostat. The temperature of the out-flowing gas/liquid from the cryostat is monitored using platinum resistor thermometer. The program allows for automatic filling at regular intervals with temperature monitoring from a remote terminal.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912040495819ZK.pdf 45KB PDF download
  文献评价指标  