期刊论文详细信息
Eurasian Journal of Analytical Chemistry
Analytical Method Development and Validation for Determination of Residual Solvent in Zopiclone Tablets by Headspace Gas Chromatography
Nainesh Lokhande1  Manish Gangrade1  Santosh Thakre1  Vishal G Shinde1  Chetan M Bhalgat1  Suraj S. Padwal1  Jayvant Harlikar1 
关键词: DB-624 column;    HSGC;    residual solvents;    validation;    zopiclone tablets;   
DOI  :  10.12973/ejac.2017.00147a
学科分类:分析化学
来源: Eurasian Journal of Analytical Chemistry
PDF
【 摘 要 】

A simple and sensitive headspace gas chromatographic method has been developed and validated for simultaneous determination of isopropyl alcohol and methylene chloride in zopiclone tablets. The separation was achieved on 75 m long DB-624 fused silica column, 0.53 mm inner diameter and 3 μ film thickness using nitrogen as a carrier gas at 6 ml/min flow and FID as a detector. The developed gas chromatographic method offers symmetric peak shape, good resolution and reasonable retention time for all the solvents. The limit of detection of isopropyl alcohol and methylene chloride was found to be 250 μg/tablet and 90 μg/tablet, respectively.