期刊论文详细信息
Bulletin of the Korean chemical society
Development and Validation of Primary Method for the Determination of Glucose in Human Serum by Isotope Dilution Liquid Chromatography Tandem Mass Spectrometry and Comparison with Field Methods
Sang Ryoul Park1  Je Hoon Lee1  Jong Man Lee1  Yong Goo Kim1  Hwa Shim Lee1 
关键词: Glucose;    ID LC/MS/MS;    Primary method;    Traceability;    Proficiency testing;   
DOI  :  
学科分类:化学(综合)
来源: Korean Chemical Society
PDF
【 摘 要 】

Glucose is a common medical analyte measuring in human serum or blood samples. The development of a primary method is necessary for the establishment of traceability in measurements. We have developed an isotope dilution liquid chromatography tandem mass spectrometry as a primary method for the measurement of glucose in human serum. Glucose and glucose-13C6 in sample were ionized in ESI negative mode and monitored at mass transfers of m/z 179/89 and 185/92 in MRM, respectively. Glucose was separated on NH2P- 50 2D column, and the mobile phase was 20 mM NH4OAc in 30% acetonitrile/70% water. Verification of this method was performed by the comparison with NIST SRMs. Our results agreed well with the SRM values. We have developed two levels of glucose serum certified reference material using this method and distributed them to the clinical laboratories in Korea as samples for proficiency testings. The expended uncertainty was about 1.2% on 95% confidence level. In proficiency testings, the results obtained from the clinical laboratories showed about 3.6% and 3.9% RSD to the certified values. Primary method can provide the traceability to the field laboratories through proficiency testings or certified reference materials.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912010244570ZK.pdf 780KB PDF download
  文献评价指标  
  下载次数:14次 浏览次数:16次