| Bulletin of the Korean chemical society | |
| Development of Multi-sample Loading Device for TEM Characterization of Hydroxyapatite Nanopowder | |
| Youn-Joong Kim1  Jung-Kyun Kim1  Eunji Lee1  Jin-Gyu Kim1  Jong-Man Jeong1  Jong-Moon Lee1  | |
| 关键词: Internal standard; Structure analysis; Multi-sample loading device; Transmission electron microscopy (TEM); Hydroxyapatite nanopowder; | |
| DOI : | |
| 学科分类:化学(综合) | |
| 来源: Korean Chemical Society | |
PDF
|
|
【 摘 要 】
A shortcoming of using transmission electron microscopy (TEM) for structural analysis via electron diffraction is the relatively large error of the measurements as compared to X-ray diffraction. To reduce these errors, various internal standard methods from earlier studies have been widely used. We developed a new device to facilitate the application of internal standard methods in preparation of TEM grids used for nanopowder analysis. Through the application of a partial mask on the TEM grid, both the internal standards and the research materials can be loaded on the same grid. Through this process, we conducted a TEM analysis that compared synthetic hydroxyapatite (HAp) nanopowder to bone apatite from a bovine femur. We determined that the accuracy of the d-spacing measurements of the HAp and bone powders could be improved to better than 1% after statistical treatments of the experimental data. By applying a quarter mask, we loaded four different nanoparticles on a single TEM grid, with one section designated for the internal standard.
【 授权许可】
Unknown
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO201912010244404ZK.pdf | 323KB |
PDF