Bulletin of the Korean chemical society | |
Performance Characteristic of the ¥â-Human Chorionic Gonadotropin Assay on the Access 2 for Its Use in Doping Control | |
Sang Mi Shin1  Changbae Jin1  Oh Seung Kwon1  Bong Chul Chung1  Hyeim Lee1  Min Jung Kang1  Hye Hyun Yoo1  | |
关键词: ¥â-hCG; Access 2; Doping control; Method evaluation; Confirmation method; | |
DOI : | |
学科分类:化学(综合) | |
来源: Korean Chemical Society | |
【 摘 要 】
The objective of this study was to validate performance characteristics of the Access 2 (Beckman coulter) system for hCG assays for use as a confirmation test for doping control. The Access 2 assay was linear up to 500 IU/L. The correlation coefficient was higher than 0.999, and the sensitivity of the linearity was 0.82. There were no false positive or false negative cases. LOD was 0.59 IU/L. The method was robust when performed by different people. Repeatability and reproducibility were below 7%. We compared reproducibility and recoveries of Access 2 and Elecsys 2010. Access 2 demonstrated higher reproducibility than Elecsys 2010. Recoveries (accuracy) of Access 2 were between 85 and 105%. Recoveries for Elecsys 2010 were between 91 and 104%.
【 授权许可】
Unknown
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201912010241869ZK.pdf | 328KB | download |