期刊论文详细信息
Bulletin of the Korean chemical society
Development of an Isotope-Dilution Flow-Injection Electrospray/Mass Spectrometric Method for the Accurate Determination of Glucosamine in Pharmaceutical Formulation
Byungjoo Kim1  Seonghee Ahn1  Yongseong Kim1  Guinam Kim1  Euijin Hwang1 
关键词: Glucosamine;    Nutritional supplement;    Flow-injection ID-ESI/MS;    Method validation;   
DOI  :  
学科分类:化学(综合)
来源: Korean Chemical Society
PDF
【 摘 要 】

An isotope-dilution flow-injection electrospray/mass spectrometric method was developed for the accurate determination of glucosamine contents in pharmaceutical formulations. Samples were extracted by methanol. After spiking glucosamine-1-13C1 as an internal standard, the extracts were then analyzed by flow-injection ESI/MS in a selected ion monitoring (SIM) mode to detect [M+H]+ ions of the analyte and its isotope analogue at m/z 180 and m/z 181, respectively. Confirmatory measurements were made by selectively monitoring the collisionally induced dissociation channels of m/z 180 �? m/z 72 and m/z 181 �?73, respectively, to test the possibility of bias in the SIM method due to matrix interferences, but any significant bias in the SIM mode was not observed. Repeatability and reproducibility studies showed that the flow-injection ESI/MS method is a reliable and reproducible method which can provide a typical method precision of 1.0 %. Other results for the method validation are reported.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912010241425ZK.pdf 307KB PDF download
  文献评价指标  
  下载次数:5次 浏览次数:4次