| Bulletin of the Korean chemical society | |
| Development of a New Copper(II) Ion-selective Poly(vinyl chloride) Membrane Electrode Based on 2-Mercaptobenzoxazole | |
| Mehrorang Ghaedi1  Javad Tashkhourian1  Morteza Akhond1  | |
| 关键词: Copper(II) ion-selective electrode; Ion sensor; PVC membrane; Potentiometry; 2-Mercaptobenzoxazole; | |
| DOI : | |
| 学科分类:化学(综合) | |
| 来源: Korean Chemical Society | |
PDF
|
|
【 摘 要 】
Copper(II) ion-selective PVC membrane electrode based on 2-mercaptobenzoxazole as a new ionophore and o-nitrophenyl octyl ether (o-NPOE) as plasticizer is proposed. This electrode revealed good selectivity for Cu2+ over a wide variety of other metal ions. Effects of experimental parameters such as membrane composition, nature and amount of plasticizer, and concentration of internal solution on the potential response of Cu2+ sensor were investigated. The electrode exhibits good response for Cu2+ in a wide linear range of 5.0 �? 10-6-1.6 �? 10-2 mol/L with a slope of 29.2 �? 2.0 mV/decade. The response time of the sensor is less than 10 s, and the detection limit is 2.0 �? 10-6 mol/L. The electrode response was stable in pH range of 4-6. The lifetime of the electrode was about 2 months. The electrode revealed comparatively good selectivities with respect to many alkali, alkaline earth, and transition metal ions.
【 授权许可】
Unknown
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO201912010239784ZK.pdf | 747KB |
PDF