Application of Methane Mixed Plasma for the Determination of Ge, As, and Se in Serum and Urine by ICP/MS
Sun-Tae Kim1  Won Lee1  Kyung-Su Park1  Young-Man Kim1  Yunje Kim1 
关键词: Methane mixed plasma;    ICP/MS;    Ge;    As;    Se;   
DOI  :  
学科分类:化学(综合)
来源: Korean Chemical Society
PDF
【 摘 要 】

An analytical method for the simultaneous determination of trace Ge, As and Se in biological samples by inductively coupled plasma/mass spectrometry has been investigated. The effects of added organic gas into the coolant argon gas on the analyte signal were studied to improve the detection limit, accuracy and precision. The addition of a small amount of methane (10 mL/min.) into the coolant gas channel improved the ionization of Ge, As and Se. The analytical sensitivity of the proposed Ar/CH4 system was superior by at least two-fold to that of the conventional Ar method. In the present method, the detection limits obtained for Ge, As and Se were 0.014, 0.012 and 0.064 mg/L, respectively. The analytical reliability of the proposed method was evaluated by analyzing the certified standard reference materials (SRM). Recoveries of 99.9% for Ge, 103% for As, 96.5% for Se were obtained for NIST SRM of freeze dried urine sample. The proposed method was also applied to the biological samples.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912010238861ZK.pdf 2247KB PDF download
  文献评价指标  
  下载次数:12次 浏览次数:28次