期刊论文详细信息
Journal of Nuclear Medicine
Clinical Performance of 2 Dedicated PET Scanners for Breast Imaging: Initial Evaluation
Mayumi Kawada1  Shotaro Kanao1  Tomoharu Sugie1  Mami Iima1  Kaori Togashi1  Takayuki Ueno1  Masakazu Toi1  Yoshiki Mikami1  Yuji Nakamoto1 
关键词: breast cancer;    dedicated;    18F-FDG;    positron emission mammography (PEM);    PET/CT;   
DOI  :  10.2967/jnumed.111.100958
学科分类:医学(综合)
来源: Society of Nuclear Medicine
PDF
【 摘 要 】

The purpose of this study was to investigate the diagnostic performance of 2 newly developed dedicated breast PET scanners in patients with known or suspected breast cancer. Methods: Two types of scanner were evaluated, an O-shaped scanner and a C-shaped scanner. The O scanner was designed for imaging patients who were prone, and the C scanner was designed for those patients positioned leaning forward. Sixty-nine women with known or suspected breast carcinoma (80 lesions: 72 invasive carcinomas, 4 noninvasive carcinomas [ductal carcinoma in situ, or DCIS], 1 case of adenomatous ductal hyperplasia, and 3 benign lesions) were enrolled in this study. All patients underwent a conventional whole-body PET/CT scan, followed by breast scanning using both dedicated devices. The diagnostic performance of each scanner was assessed. Results: The maximal diameter of invasive tumors ranged from 4 to 112 mm, with an average of 26 mm. With the O scanner, 62 of 76 malignant lesions (including 3 DCIS) were detected, 5 lesions were not detected, and the remaining 9 lesions were outside the field of view. With the C scanner, 63 of 76 malignant lesions (including 2 DCIS) were detected, 7 lesions were not detected, and the remaining 6 lesions were outside the field of view. The lesion-based sensitivities of the O and C scanners were 82% (62/76) and 83% (63/76), respectively; sensitivities excluding lesions outside the field of view were 93% (62/67) and 90% (63/70), respectively. The sensitivity of conventional PET/CT was 92% (70/76). All lesions outside the field of view were close to the chest wall. The breast-based specificities of the O, C, and conventional scanners were 98% (48/49), 98% (56/57), and 100% (70/70), respectively. Conclusion: Our preliminary study indicates that both dedicated breast PET scanners are clinically feasible and yield reasonably high sensitivity. More detailed information was obtained with these scanners than with the conventional scanner.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912010198264ZK.pdf 1012KB PDF download
  文献评价指标  
  下载次数:9次 浏览次数:17次