| Journal of Nuclear Medicine | |
| Simple Analytic Method of 11C-Flumazenil Metabolite in Blood | |
| Koichi Ishizu1  Junji Konishi1  Masafumi Ihara1  Yasuhiro Magata1  Hideo Saji1  Sadahiko Nishizawa1  Haruhiro Kitano1  Takahiro Mukai1  | |
| 关键词: positron; benzodiazemine receptor; flumazenil; input function; | |
| DOI : | |
| 学科分类:医学(综合) | |
| 来源: Society of Nuclear Medicine | |
PDF
|
|
【 摘 要 】
11C-Flumazenil (11C-FMZ) is useful to estimate central benzodiazepine receptors by PET. The binding potential (BP) can be calculated with dynamic PET and continual blood sampling. Because conventional metabolite analysis of plasma samples is complicated and time consuming, a simple method is required to obtain an input function. In this article, a whole blood solvent extraction method was evaluated using data of 13 subjects. Methods: The plasma solvent extraction method was estimated in comparison with the thin-layer chromatography (TLC) method. Then, the whole blood solvent extraction method was evaluated in comparison with the plasma solvent extraction method. Results: Metabolite data analyzed by the plasma extraction method were well correlated with those by the TLC method (r = 0.99). The BP was calculated using both the whole blood extraction data and the plasma extraction data. No difference was observed in all brain regions. Conclusion: This simple whole blood solvent extraction method can be applied to clinical BP estimation using 11C-FMZ.
【 授权许可】
Unknown
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO201912010195510ZK.pdf | 574KB |
PDF