[1] University of Oklahoma and Department of Veterans Affairs Medical Center, Oklahoma City, Oklahoma and Sun Nuclear Corporation, Melbourne, Florida University of Oklahoma and Department of Veterans Affairs Medical Center, Oklahoma City, Oklahoma and Sun Nuclear Corporation, Melbourne, Florida University of Oklahoma and Department of Veterans Affairs Medical Center, Oklahoma City, Oklahoma and Sun Nuclear Corporation, Melbourne, Florida
关键词: dose calibrator;    accuracy;    constancy;    linearity;    geometry;    quality control;   
DOI  :  
学科分类:医学(综合)
来源: Society of Nuclear Medicine
PDF
【 摘 要 】

We have reviewed quality control testing of dose calibrators and have analyzed the potential for error. The uncertainty in half-lives of reference sources can lead to inaccurate assessment of the accuracy and constancy in dose calibrator performance. Molybdenum-99 contamination of 99mTc used as a reference source can appear as a nonlinear response of the dose calibrator. The useful life of a reference source and the tolerable amount of 99Mo contamination are examined.