期刊论文详细信息
Journal of Veterinary Medical Science
Electroretinography Using Contact Lens Electrode with Built-In Light Source in Dogs
Keiko TSUZUKI1  Tadao KOTANI1  Yoshiki ITOH1  Shinsuke WAKAIKI1  Takahiro SENO1  Seiya MAEHARA1  Norihiko ITOH1  Kazuto YAMASHITA1  Tokiko KUSHIRO1  Yasuharu IZUMISAWA1 
[1] Department of Companion Animal Clinical Sciences, School of Veterinary Medicine, Rakuno Gakuen University
关键词: canine;    contact lens electrode;    electroretinography;    rod response;    standard flash;   
DOI  :  10.1292/jvms.67.509
学科分类:兽医学
来源: Japanese Society of Veterinary Science
PDF
【 摘 要 】

References(25)Cited-By(16)Electroretinography (ERG) is an effective method for the diagnosis of retinal disease. In the dog, dependable ERG recording is difficult without the use of an expensive device like a Ganzfeld full-field stimulator. The International Society for Clinical Electrophysiology of Vision has defined the standard flash stimulus condition (SF) and evaluation of the retina using the b/a ratio in humans. In dogs, evaluation using the b/a ratio has not been reported, whereas the intensity of SF has been defined. In this study, we performed a convenient ERG recording method using a contact lens electrode with a built-in light source (LED-electrode), and confirmed SF as reported previously. ERG recordings were performed on 15 healthy beagle dogs under sedation. We performed bilateral ERG at 12 different intensities after 30 min dark adaptation. After 10 min light adaptation, we recorded single flash cone and flicker cone response using the SF determined in this study. In this study, SF of 3.0 cd/m2/sec (6,000 cd/m2, 0.5 msec) resulted in b/a=2. The intensity for rod response that recorded only the b-wave was 0.0096 cd/m2/sec (80 cd/m2, 0.12 msec). We could achieve ERG for each response easily and smoothly under sedation, and without general anesthesia. Using an LED-electrode, we could perform more quantitative and reproducible ERG examinations than with traditional methods. We propose that the b/a ratio is the most useful parameter in ERG reporting for evaluating retinal function.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201911300821129ZK.pdf 365KB PDF download
  文献评价指标  
  下载次数:7次 浏览次数:9次