Comparison of Two Electroretinography Systems Used in Dogs: The HMsERG and the RETIport
Shin Ae PARK2  Kang Moon SEO1  Se Eun KIM1  Won Gyun SON1  Young Woo PARK1  Man Bok JEONG1 
[1] Department of Veterinary Surgery and Ophthalmology, College of Veterinary Medicine and BK21 Program for Veterinary Science, Seoul National University;Department of Surgical and Radiological Sciences, School of Veterinary Medicine, University of California Davis
关键词: dog standard protocol;    electroretinography;    HMsERG;    retinal function;    RETIport;   
DOI  :  10.1292/jvms.10-0197
学科分类:兽医学
来源: Japanese Society of Veterinary Science
PDF
【 摘 要 】

References(35)The purpose of the study was to compare two different electroretinography (ERG) instruments used on the same animal in a laboratory setting. Retinal function in both eyes of 12 healthy miniature schnauzers was evaluated under general anesthesia. Scotopic and photopic ERGs were recorded by the HMsERG and the RETIport using the Dog Standard Protocol recommended by the European College of Veterinary Ophthalmologists (ECVO). The waveforms of the ERGs obtained by the two ERG units were similar to those described in previous studies. The 90% reference ranges using the multiple of medians (MoM) showed that the changes in ERG parameters obtained by the HMsERG unit were very similar to those of the RETIport for most ERG responses, except for a few. The results demonstrate that the two ERG systems are comparable for evaluating retinal function in dogs. Further, the results also show that it is necessary to establish ERG system-specific limits of normality in each laboratory or clinic in order to obtain a correct diagnosis, for example by using a graphical representation of the 90% reference range.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201911300306075ZK.pdf 950KB PDF download
  文献评价指标  
  下载次数:8次 浏览次数:6次