| IEICE Electronics Express | |
| RF power measurement in D-band using down-converter calibrated by three-mixer method | |
| Katsumi Fujii1  Yasushi Matsumoto1  Kaori Fukunaga1  Toshihide Tosaka1  | |
| [1] National Institute of Information and Communications Technology | |
| 关键词: D-band; reciprocal mixer; RF power standard; conversion loss; | |
| DOI : 10.1587/elex.9.1096 | |
| 学科分类:电子、光学、磁材料 | |
| 来源: Denshi Jouhou Tsuushin Gakkai | |
PDF
|
|
【 摘 要 】
References(2)Cited-By(1)To establish an RF power standard above 110GHz, we have developed a measurement system using a down-conversion mixer with a spectrum analyzer, and the conversion loss of the mixer was determined by the three-mixer method. This method measures the three transmission S-parameters (S21) of three mixer pairs using a vector network analyzer. The experimental results proved that the three-mixer method can determine the conversion loss of mixers with an uncertainty of 0.5dB (level of confidence: 95%) in the entire range of the D-band. Using this uncertainty and the uncertainty of the spectrum analyzer, the proposed system can measure the RF power in the D-band with an uncertainty of less than 2.0dB.
【 授权许可】
Unknown
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO201911300784622ZK.pdf | 452KB |
PDF