Journal of Veterinary Medical Science | |
Indirect arterial blood pressure measurement in healthy anesthetized catsusing a device that combines oscillometry with photoplethysmography | |
Naoyuki ITOH1  Fumio HOSHI1  Yasutomo HORI1  Yasuhiro HEISHIMA2  Seishiro CHIKAZAWA1  Kazutaka KANAI1  | |
[1] Laboratory of Small Animal Internal Medicine II, School of Veterinary Medicine, Kitasato University, 23�?35�?1 Higashi, Towada, Aomori 034�?8628, Japan;Heisei Animal Hospital, 2�?1�?1 Futagocho, Kasugai, Aichi 486�?0955, Japan | |
关键词: blood pressure; feline; oscillometry; photoplethysmography; systemic hypertension; | |
DOI : 10.1292/jvms.15-0612 | |
学科分类:兽医学 | |
来源: Japanese Society of Veterinary Science | |
【 摘 要 】
References(14)We investigated the basic characteristics of indirect arterial blood pressure (ABP)measurement using a device that combines oscillometry and photoplethysmography in cats.Dobutamine was infused intravenously in four anesthetized cats. Direct ABP was measured bya catheter. Indirect ABP was measured from the left forelimb. Dobutamine significantlyelevated both systolic arterial pressure (SAP) and mean arterial pressure (MAP) in adose-dependent manner. The indirect SAP, MAP and diastolic arterial pressure (DAP) valueswere closely correlated with the direct ABP values (r=0.88, 0.89 and0.83, respectively). The mean bias for SAP, MAP and DAP was 3.4, 0.2 and �?2.4 mmHg,respectively. The indirect ABP measured by this device may be used to reliably monitor ABPchanges in anesthetized cats.
【 授权许可】
Unknown
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201911300157141ZK.pdf | 537KB | download |