期刊论文详细信息
卷:173
Rapid screening of mycotoxins in liquid milk and milk powder by automated size-exclusion SPE-UPLC-MS/MS and quantification of matrix effects over the whole chromatographic run
Wang, Xiupin ; Li, Peiwu
关键词: Liquid milk;    Milk powder;    UPLC-MS/MS;    Automated exclusion SPE;    Matrix effects;    Permanent postcolumn infusion (PCI);   
DOI  :  10.1016/j.foodchem.2014.10.056
学科分类:食品科学和技术
PDF
【 摘 要 】

An automated, size-exclusion solid phase extraction (SPE)-UPLC-MS/MS protocol without pre-treatment of samples was developed to screen for four mycotoxins (OTA, ZEN, AFB(1), and AFM(1)) in liquid milk and milk powder. Firstly, a mixed macropore-silica gel cartridge was established as a size-exclusion SPE column. The proposed methodology could be a candidate in green analytical chemistry because it saves on manpower and organic solvent. Permanent post-column infusion of mycotoxin standards was used to quantify matrix effects throughout the chromatographic run. Matrix-matched calibration could effectively compensate for matrix effects, which may be caused by liquid milk or milk powder matrix. Recovery of the four mycotoxins in fortified liquid milk was in the range 89-120% and RSD 2-9%. The LOD for the four mycotoxins in liquid milk and milk powder were 0.05-2 ng L-1 and 0.25-10 ng kg(-1), respectively. The LOQ for the four mycotoxins in liquid milk and milk powder were 0.1-5 ng L-1 and 0.5-25 ng kg(-1), respectively. (C) 2014 Elsevier Ltd. All rights reserved.

【 授权许可】

   

【 预 览 】
附件列表
Files Size Format View
JA201706070005563SK.pdf KB PDF download
  文献评价指标  
  下载次数:3次 浏览次数:5次