Peripheral doses in patients undergoing Cyberknife treatment for intracranial lesions. A single centre experience | |
George Panayiotakis2  Dimitrios Kardamakis3  Nikolaos Salvaras1  Argyrios Tzouras1  Harry Delis2  Christos Antypas1  Vassiliki Vlachopoulou2  | |
[1] CyberKnife Center, Iatropolis, Athens, Attica, Greece;Department of Medical Physics, School of Medicine, University of Patras, Achaia, Greece;Department of Radiology, School of Medicine, University Hospital of Patras, Achaia, Greece | |
关键词: stereotactic/radiosurgery radiotherapy; peripheral dose; MOSFET; intracranial lesions; Cyberknife; | |
Others : 1223816 DOI : 10.1186/1748-717X-6-157 |
|
received in 2011-09-04, accepted in 2011-11-14, 发布年份 2011 | |
【 摘 要 】
Background
Stereotactic radiosurgery/radiotherapy procedures are known to deliver a very high dose per fraction, and thus, the corresponding peripheral dose could be a limiting factor for the long term surviving patients. The aim of this clinical study was to measure the peripheral dose delivered to patients undergoing intracranial Cyberknife treatment, using the MOSFET dosimeters. The influence of the supplemental shielding, the number of monitor units and the collimator size to the peripheral dose were investigated.
Methods
MOSFET dosimeters were placed in preselected anatomical regions of the patient undergoing Cyberknife treatment, namely the thyroid gland, the nipple, the umbilicus and the pubic symphysis.
Results
The mean peripheral doses before the supplemental shielding was added to the Cyberknife unit were 51.79 cGy, 13.31 cGy and 10.07 cGy while after the shielding upgrade they were 38.40 cGy, 10.94 cGy, and 8.69 cGy, in the thyroid gland, the umbilicus and the pubic symphysis, respectively. The increase of the collimator size corresponds to an increase of the PD and becomes less significant at larger distances, indicating that at these distances the PD is predominate due to the head leakage and collimator scatter.
Conclusion
Weighting the effect of the number of monitor units and the collimator size can be effectively used during the optimization procedure in order to choose the most suitable treatment plan that will deliver the maximum dose to the tumor, while being compatible with the dose constraints for the surrounding organs at risk. Attention is required in defining the thyroid gland as a structure of avoidance in the treatment plan especially in patients with benign diseases.
【 授权许可】
2011 Vlachopoulou et al; licensee BioMed Central Ltd.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
20150905081805131.pdf | 1850KB | download | |
Figure 8. | 18KB | Image | download |
Figure 7. | 15KB | Image | download |
Figure 6. | 23KB | Image | download |
Figure 5. | 19KB | Image | download |
Figure 4. | 19KB | Image | download |
Figure 3. | 17KB | Image | download |
Figure 2. | 19KB | Image | download |
Figure 1. | 99KB | Image | download |
【 图 表 】
Figure 1.
Figure 2.
Figure 3.
Figure 4.
Figure 5.
Figure 6.
Figure 7.
Figure 8.
【 参考文献 】
- [1]Xu XG, Bednarz B, Paganetti H: A review of dosimetry studies on external-beam radiation treatment with respect to second cancer induction. Phys Med Biol 2008, 53(13):R193-R241.
- [2]Brenner DJ, Doll R, Goodhead DT, Hall EJ, Land CE, Little JB, Lubin JH, Preston DL, Preston RJ, Puskin JS, Ron E, Sachs RK, Samet JM, Setlow RB, Zaider M: Cancer risks attributable to low doses of ionizing radiation: Assessing what we really know. Proceedings of the National Academy of Sciences of the United States of America 2003, 100(24):13761-13766.
- [3]Acun H, Kemikler G, Karadeniz A: Dosimetric analysis of thyroid doses from total cranial irradiation. Radiat Prot Dosimetry 2007, 123(4):498-504.
- [4]Somerville HM, Steinbeck KS, Stevens G, Delbridge LW, Lam AH, Stevens MM: Thyroid neoplasia following irradiation in adolescent and young adult survivors of childhood cancer. Med J Aust 2002, 176(12):584-587.
- [5]Antypas C, Pantelis Ε: Performance evaluation of a CyberKnife® G4 image-guided robotic stereotactic radiosurgery system. Phys Med Biol 2008, 53:4697-4718.
- [6]Murphy MJ: Fiducial-based targeting accuracy for external-beam radiotherapy. Med Phys 2002, 29(3):334-344.
- [7]Fu D, Kuduvalli G, Mitrovic V, Main W, Thomson L: Automated skull tracking for the CyberKnife® image-guided radiosurgery system. Progress in Biomedical Optics and Imaging - Proceedings of SPIE 2005, 5744(1):366-377.
- [8]Ho AK, Fu D, Cotrutz C, Hancock SL, Chang SD, Gibbs IC, Maurer CR Jr, Adler JR Jr: A study of the accuracy of cyberknife spinal radiosurgery using skeletal structure tracking. Neurosurgery 2007, 60(2 Suppl 1):147-156.
- [9]Collins SP, Coppa ND, Zhang Y, Collins BT, McRae DA, Jean WC: CyberKnife® radiosurgery in the treatment of complex skull base tumors: analysis of treatment planning parameters. Radiat Oncol 2006, 1(1):46. BioMed Central Full Text
- [10]Zytkovicz A, Daftari I, Phillips TL, Chuang CF, Verhey L, Petti PL: Peripheral dose in ocular treatments with CyberKnife and Gamma Knife radiosurgery compared to proton radiotherapy. Phys Med Biol 2007, 52(19):5957-5971.
- [11]Petti PL, Chuang CF, Smith V, Larson DA: Peripheral doses in Cyberknife radiosurgery. Med Phys 2006, 33(6):1770-1779.
- [12]Chuang CF, Larson DA, Zytkovicz A, Smith V, Petti PL: Peripheral dose measurement for CyberKnife radiosurgery with upgraded linac shielding. Med Phys 2008, 35(4):1494-1496.
- [13]Di Betta E, Fariselli L, Bergantin A, Locatelli F, Del Vecchio A, Broggi S, Fumagalli ML: Evaluation of the peripheral dose in stereotactic radiotherapy and radiosurgery treatments. Med Phys 2010, 37(7):3587-3594.
- [14]Yu C, Luxton G, Apuzzo ML, MacPherson DM, Petrovich Z: Extracranial radiation doses in patients undergoing gamma knife radiosurgery. Neurosurgery 1997, 41(3):553-559.
- [15]Vlachopoulou V, Malatara G, Delis H, Theodorou K, Kardamakis D, Panayiotakis G: Peripheral dose measurement in high-energy photon radiotherapy with the implementation of MOSFET. World J Radiol 2010, 2(11):434-439.
- [16]Consorti R, Petrucci A, Fortunato F, Soriani A, Marzi S, Iaccarino G, Landoni V, Benassi M: In vivo dosimetry with MOSFETs: Dosimetric characterization and first clinical results in intraoperative radiotherapy. Int J Radiat Oncol Biol Phys 2005, 63(3):952-960.
- [17]Rosenfeld AB: MOSFET Dosimetry on Modern Radiation Oncology Modalities. Radiat Prot Dosimetry 2002, 101:393-398.
- [18]Cheung T, Yu PKN, Butson MJ: Low-dose measurement with a MOSFET in high-energy radiotherapy applications. Radiat Meas 2005, 39(1):91-94.
- [19]Ramani R, Russell S, O'Brien P: Clinical dosimetry using mosfets. Int J Radiat Oncol Biol Phys 1997, 37(4):959-964.
- [20]Butson MJ, Cheung T, Yu PKN: Peripheral dose measurement with a MOSFET detector. Appl Radiat Isotopes 2005, 62(4):631-634.
- [21]Butson MJ, Rozenfeld A, Mathur JN, Carolan M, Wong TP, Metcalfe PE: A new radiotherapy surface dose detector: the MOSFET. Med Phys 1996, 23(5):655-658.
- [22]Best Medical Canada: Operator's manual for the mobile MOSFET dosimetry system. Ottawa; 2007.
- [23]Kilby W, Dooley JR, Kuduvalli G, Sayeh S, Maurer CR Jr: The CyberKnife® Robotic Radiosurgery System in 2010. Technol Cancer Res Treat 2010, 9(5):433-552.
- [24]Stovall M, Blackwell CR, Cundiff J, Novack DH, Palta JR, Wagner LK, Webster EW, Shalek RJ: Fetal Dose from radiotherapy with photon beams: Report of AAPM Radiation Therapy Committee Task Group N° 36. Med Phys 1995, 22(1):63-82.
- [25]Antypas C, Sandilos P, Kouvaris J, Balafouta E, Karinou E, Kollaros N, Vlahos L: Fetal Dose Evaluation During Breast Cancer Radiotherapy. Int J Radiat Oncol Biol Phys 1998, 40(4):995-999.
- [26]International Commission of Radiation Protection (ICRP): The 2007 Recommendations of the International Commission on Radiological Protection. Oxford: Pergamon; 2007.