Target splitting non-coplanar RapidArc radiation therapy for a diffuse sebaceous carcinoma of the scalp: a novel delivery technique
ZhenYu Qi1  ALong Chen1  DeHua Kang1  ZhiChun He1  WeiWei Xiao1  Jiang Hu1 
[1] Department of Radiation Oncology, Sun Yat-Sen University Cancer Center, Guangzhou 510060, China
关键词: Dosimetry;    Non-coplanar;    Target splitting;    RapidArc;    Total scalp irradiation;   
Others  :  1151372
DOI  :  10.1186/1748-717X-9-204
 received in 2013-11-10, accepted in 2014-08-28,  发布年份 2014
PDF
【 摘 要 】

Background and purpose

To compare conventional lateral photon-electron, fixed-beam intensity modulated radiation therapy (IMRT), coplanar and non-coplanar RapidArc for the treatment of a diffuse sebaceous gland carcinoma of the scalp.

Methods

Comprehensive dosimetry comparisons were performed among 3D-CRT, IMRT and various RapidArc plans. Target coverage, conformity index (CI), homogeneity index (HI) and doses to organs at risk (OAR) were calculated. Monitor unites (MUs) and delivery time of each treatment were also recorded to evaluate the execution efficiency. The influence of target splitting technique and non-coplanar planning on plan quality was discussed.

Results

IMRT was superior to 3D-CRT concerning targets’ coverage at the sacrifice of larger irradiated brain volumes to low doses. CIs and HIs were better in coplanar RapidArc and non-coplanar RapidArc plans than 3D-CRT and IMRT. Best dose coverage and sparing of OARs were achieved in non-coplanar plans using target splitting technique. Treatment delivery time was longest in the IMRT plan and shortest in the coplanar RapidArc plan without target splitting. The 3%/3 mm gamma test pass rates were above 95% for all the plans.

Conclusions

Target splitting technique and non-coplanar arcs are recommended for total scalp irradiation.

【 授权许可】

   
2014 Hu et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150406065553672.pdf 2972KB PDF download
Figure 3. 55KB Image download
Figure 2. 96KB Image download
Figure 1. 86KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

【 参考文献 】
  • [1]Rizvi M, De Jesus R, Girotto J, Nonaka D, Goldberg N: Sebaceous gland carcinoma of the scalp: case report and review of the literature. J Otolaryngol 2003, 32:64-68.
  • [2]Hata M, Koike I, Omura M, Maegawa J, Ogino I, Inoue T: Noninvasive and curative radiation therapy for sebaceous carcinoma of the eyelid. Int J Radiat Oncol Biol Phys 2012, 82:605-611.
  • [3]Able CM, Mills MD, McNeese MD, Hogstrom KR: Evaluation of a total scalp electron irradiation technique. Int J Radiat Oncol Biol Phys 1991, 21:1063-1072.
  • [4]Mellenberg DE, Schoeppel SL: Total scalp treatment of mycosis fungoides: the 4 x 4 technique. Int J Radiat Oncol Biol Phys 1993, 27:953-958.
  • [5]Akazawa C: Treatment of the scalp using photon and electron beams. Med Dosim 1989, 14:129-131.
  • [6]Wojcicka JB, Lasher DE, McAfee SS, Fortier GA: Dosimetric comparison of three different treatment techniques in extensive scalp lesion irradiation. Radiother Oncol 2009, 91:255-260.
  • [7]Kelly PJ, Mannarino E, Lewis JH, Baldini EH, Hacker FL: Total dural irradiation: RapidArc versus static-field IMRT: a case study. Med Dosim 2012, 37:175-181.
  • [8]International Commission on Radiation Units and Measurements: Report 50: Prescribing, recording, and reporting photon beam therapy. Bethesda, MD: ICRU; 1993.
  • [9]International Commission on Radiation Units and Measurements: Report 62: Prescribing, recording and reporting photon beam therapy (supplement to ICRU report 50). Bethesda, MD: ICRU; 1999.
  • [10]Feuvret L, Noël G, Mazeron JJ, Bey P: Conformity index: a Review. Int J Radiat Oncol Biol Phys 2006, 64:333-342.
  • [11]Kataria T, Sharma K, Subramani V, Karrthick KP, Bisht SS: Homogeneity Index: An objective tool for assessment of conformal radiation treatments. J Med Phys 2012, 37:207-213.
  • [12]Tung SS, Shiu AS, Starkschall G, Morrison WH, Hogstrom KR: Dosimetric evaluation of total scalp irradiation using a lateral electron-photon technique. Int J Radiat Oncol Biol Phys 1993, 27:153-160.
  • [13]Guckenberger M, Kavanagh A, Partridge M: Combining advanced radiotherapy technologies to maximize safety and tumor control probability in stage III non-small cell lung cancer. Strahlenther Onkol 2012, 188:894-900.
  • [14]Mounessi FS, Lehrich P, Haverkamp U, Willich N, Bölling T, Eich HT: Pelvic Ewing sarcomas. Three-dimensional conformal vs. intensity-modulated radiotherapy. Strahlenther Onkol 2013, 189:308-314.
  • [15]Bedford JL, Childs PJ, Hansen VN, Warrington AP, Mendes RL, Glees JP: Treatment of extensive scalp lesions with segmental intensity-modulated photon therapy. Int J Radiat Oncol Biol Phys 2005, 62:1549-1558.
  • [16]Sahgal A, Chuang C, Larson D, Huang K, Petti P, Weinstein P, Ma L: Split-volume treatment planning of multiple consecutive vertebral body metastases for cyberknife image-guided robotic radiosurgery. Med Dosim 2008, 33:175-179.
  • [17]Seppälä J, Kulmala J, Lindholm P, Minn H: A method to improve target dose homogeneity of craniospinal irradiation using dynamic split field IMRT. Radiother Oncol 2010, 96:209-215.
  • [18]Wurstbauer K, Deutschmann H, Kopp P, Merz F, Schöller H, Sedlmayer F: Target splitting in radiation therapy for lung cancer: further developments and exemplary treatment plans. Radiat Oncol 2009, 4:30. BioMed Central Full Text
  • [19]Wurstbauer K, Deutschmann H, Kranzinger M, Merz F, Rahim H, Sedlmayer F, Kogelnik HD: Radiotherapy for lung cancer: Target splitting by asymmetric collimation enables reduction of radiation doses to normal tissues and dose escalation. Int J Radiat Oncol Biol Phys 1999, 44:333-341.
  • [20]Laing RW, Bentley RE, Nahum AE, Warrington AP, Brada M: Stereotactic radiotherapy of irregular targets: a comparison between static conformal beams and non-coplanar arcs. Radiother Oncol 1993, 28:241-246.
  • [21]Krayenbuehl J, Davis JB, Ciernik IF: Dynamic intensity-modulated non-coplanar arc radiotherapy (INCA) for head and neck cancer. Radiother Oncol 2006, 81:151-157.
  • [22]Ward JW, Phillips R, Williams T, Shang C, Page L, Prest C, Beavis AW: Immersive visualization with automated collision detection for radiotherapy treatment planning. Stud Health Technol Inform 2007, 125:491-496.
  • [23]Dong P, Lee P, Ruan D, Long T, Romeijn E, Yang Y, Low D, Kupelian P, Sheng K: 4π non-coplanar liver SBRT: a novel delivery technique. Int J Radiat Oncol Biol Phys 2013, 85:1360-1366.
  • [24]Soisson ET, Tomé WA, Richards GM, Mehta MP: Comparison of linac based fractionated stereotactic radiotherapy and tomotherapy treatment plans for skull-base tumors. Radiother Oncol 2006, 78:313-321.
  文献评价指标  
  下载次数:82次 浏览次数:24次