Radiation Oncology
Dual-gated volumetric modulated arc therapy
Lei Xing1  Sarah Geneser1  Huanmei Wu2  Junqing Wu3  Benjamin Fahimian1 
[1] Department of Radiation Oncology, Stanford University, Stanford, CA, USA;Purdue School of Engineering and Technology, Indiana University School of Informatics, IUPUI, Indianapolis, IN, USA;School of Health Sciences, Purdue University, West Lafayette, IN, USA
关键词: SBRT;    VMAT;    IGRT;    Dual gating;   
Others  :  1151273
DOI  :  10.1186/1748-717X-9-209
 received in 2013-10-30, accepted in 2014-08-29,  发布年份 2014
PDF
【 摘 要 】

Background

Gated Volumetric Modulated Arc Therapy (VMAT) is an emerging radiation therapy modality for treatment of tumors affected by respiratory motion. However, gating significantly prolongs the treatment time, as delivery is only activated during a single respiratory phase. To enhance the efficiency of gated VMAT delivery, a novel dual-gated VMAT (DG-VMAT) technique, in which delivery is executed at both exhale and inhale phases in a given arc rotation, is developed and experimentally evaluated.

Methods

Arc delivery at two phases is realized by sequentially interleaving control points consisting of MUs, MLC sequences, and angles of VMAT plans generated at the exhale and inhale phases. Dual-gated delivery is initiated when a respiration gating signal enters the exhale window; when the exhale delivery concludes, the beam turns off and the gantry rolls back to the starting position for the inhale window. The process is then repeated until both inhale and exhale arcs are fully delivered. DG-VMAT plan delivery accuracy was assessed using a pinpoint chamber and diode array phantom undergoing programmed motion.

Results

DG-VMAT delivery was experimentally implemented through custom XML scripting in Varian’s TrueBeam™ STx Developer Mode. Relative to single gated delivery at exhale, the treatment time was improved by 95.5% for a sinusoidal breathing pattern. The pinpoint chamber dose measurement agreed with the calculated dose within 0.7%. For the DG-VMAT delivery, 97.5% of the diode array measurements passed the 3%/3 mm gamma criterion.

Conclusions

The feasibility of DG-VMAT delivery scheme has been experimentally demonstrated for the first time. By leveraging the stability and natural pauses that occur at end-inspiration and end-exhalation, DG-VMAT provides a practical method for enhancing gated delivery efficiency by up to a factor of two.

【 授权许可】

   
2014 Fahimian et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150406052409705.pdf 3780KB PDF download
Figure 4. 65KB Image download
Figure 3. 165KB Image download
Figure 2. 112KB Image download
Figure 1. 131KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

Figure 4.

【 参考文献 】
  • [1]Mah D, Hanley J, Rosenzweig KE, Yorke E, Braban L, Ling CC, Leibel SA, Mageras G: Technical aspects of the deep inspiration breath-hold technique in the treatment of thoracic cancer. Int J Radiat Oncol Biol Phys 2000, 48:1175-1185.
  • [2]Langen KM, Jones DT: Organ motion and its management. Int J Radiat Oncol Biol Phys 2001, 50:265-278.
  • [3]Xing L, Thorndyke B, Schreibmann E, Yang Y, Li TF, Kim GY, Luxton G, Koong A: Overview of image-guided radiation therapy. Med Dosim 2006, 31:91-112.
  • [4]Shirato H, Onimaru R, Ishikawa M, Kaneko J, Takeshima T, Mochizuki K, Shimizu S, Umegaki K: Real-time 4-D radiotherapy for lung cancer. Cancer Sci 2012, 103:1-6.
  • [5]Sharp GC, Jiang SB, Shimizu S, Shirato H: Prediction of respiratory tumour motion for real-time image-guided radiotherapy. Phys Med Biol 2004, 49:425-440.
  • [6]Kubo HD, Len PM, Minohara S, Mostafavi H: Breathing-synchronized radiotherapy program at the University of California Davis Cancer Center. Med Phys 2000, 27:346-353.
  • [7]Pepin EW, Wu H, Shirato H: Dynamic gating window for compensation of baseline shift in respiratory-gated radiation therapy. Med Phys 2011, 38:1912-1918.
  • [8]Mihaylov IB, Fatyga M, Moros EG, Penagaricano J, Lerma FA: Lung dose for minimally moving thoracic lesions treated with respiration gating. Int J Radiat Oncol Biol Phys 2010, 77:285-291.
  • [9]Berbeco RI, Nishioka S, Shirato H, Jiang SB: Residual motion of lung tumors in end-of-inhale respiratory gated radiotherapy based on external surrogates. Med Phys 2006, 33:4149-4156.
  • [10]Wu J, Li H, Shekhar R, Suntharalingam M, D'Souza W: An evaluation of planning techniques for stereotactic body radiation therapy in lung tumors. Radiother Oncol 2008, 87:35-43.
  • [11]Wink NM, Chao M, Antony J, Xing L: Individualized gating windows based on four-dimensional CT information for respiration-gated radiotherapy. Phys Med Biol 2008, 53:165-175.
  • [12]Otto K: Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys 2008, 35:310-317.
  • [13]Crooks SM, Wu X, Takita C, Matzich M, Xing L: Aperture modulated arc therapy. Phys Med Biol 2003, 48:1333-1344.
  • [14]Yu CX: Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapy. Phys Med Biol 1995, 40:1435-1449.
  • [15]Fox T, Simon EL, Elder E, Riffenburgh RH, Johnstone PA: Free breathing gated delivery (FBGD) of lung radiation therapy: analysis of factors affecting clinical patient throughput. Lung Cancer 2007, 56:69-75.
  • [16]Murphy MJ, Lin PS, Ozhasoglu C: Intra-fraction dose delivery timing during stereotactic radiotherapy can influence the radiobiological effect. Med Phys 2007, 34:481-484.
  • [17]Chin E, Otto K: Investigation of a novel algorithm for true 4D-VMAT planning with comparison to tracked, gated and static delivery. Med Phys 2011, 38:2698-2707.
  • [18]Geneser S, Kirby N, Xing L, Pouliot J: The Impact of Errors in Non-rigid Medical Image Registration Methods on Dose Warping Accuracy. 2012. [ASTRO Proceedings]
  • [19]Geneser S, Fahimian B, Xing L: Improving gated delivery efficiency using brief breath holds at both inhale and exhale. Med Phys 2013, 40:296.
  • [20]Suh Y, Sawant A, Venkat R, Keall PJ: Four-dimensional IMRT treatment planning using a DMLC motion-tracking algorithm. Phys Med Biol 2009, 54:3821-3835.
  • [21]Trofimov A, Rietzel E, Lu HM, Martin B, Jiang S, Chen GT, Bortfeld T: Temporo-spatial IMRT optimization: concepts, implementation and initial results. Phys Med Biol 2005, 50:2779-2798.
  • [22]Suh Y, Weiss E, Zhong H, Fatyga M, Siebers JV, Keall PJ: A deliverable four-dimensional intensity-modulated radiation therapy-planning method for dynamic multileaf collimator tumor tracking delivery. Int J Radiat Oncol Biol Phys 2008, 71:1526-1536.
  • [23]Gong G, Wang R, Guo Y, Zhai D, Liu T, Lu J, Chen J, Liu C, Yin Y: Reduced lung dose during radiotherapy for thoracic esophageal carcinoma: VMAT combined with active breathing control for moderate DIBH. Radiat Oncol 2013, 8:291. BioMed Central Full Text
  • [24]Thomas E, Popple RA, Prendergast BM, Clark GM, Dobelbower MC, Fiveash JB: Effects of flattening filter-free and volumetric-modulated arc therapy delivery on treatment efficiency. J Appl Clin Med Phys 2013, 14:6.
  文献评价指标  
  下载次数:11次 浏览次数:3次