Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine | |
Mechanical versus manual chest compressions for cardiac arrest: a systematic review and meta-analysis | |
Xiaoli Jing2  Xiang Zhao2  Yi Yu2  Dongping Wang1  Hui Li2  | |
[1] Department of Organ Transplantation, The First Affiliated Hospital of Sun Yat-Sen University, Guangzhou 510080, China;Department of Emergency, The First Affiliated Hospital of Sun Yat-Sen University, 58 Zhongshan 2nd Road, Guangzhou 510080, Guangdong, China | |
关键词: Meta-analysis; Device; Cardiopulmonary resuscitation; Cardiac arrest; | |
Others : 1235803 DOI : 10.1186/s13049-016-0202-y |
|
received in 2015-11-12, accepted in 2016-01-21, 发布年份 2016 | |
【 摘 要 】
Background
The aim of this paper was to conduct a systematic review of the published literatures comparing the use of mechanical chest compression device and manual chest compression during cardiac arrest (CA) with respect to short-term survival outcomes and neurological function.
Methods
Databases including MEDLINE, EMBASE, Web of Science and the ClinicalTrials.gov registry were systematically searched. Further references were gathered from cross-references from articles by handsearch. The inclusion criteria for this review must be human prospective controlled studies of adult CA. Random effects models were used to assess the risk ratios and 95 % confidence intervals for return of spontaneous circulation (ROSC), survival to admission and discharge, and neurological function.
Results
Twelve trials (9 out-of-hospital and 3 in-hospital studies), involving 11,162 participants, were included in the review. The results of this meta-analysis indicated no differences were found in Cerebral Performance Category (CPC) scores, survival to hospital admission and survival to discharge between manual cardiopulmonary resuscitation (CPR) and mechanical CPR for out-of-hospital CA (OHCA) patients. The data on achieving ROSC in both of in-hospital and out-of-hospital setting suggested poor application of the mechanical device (RR 0.71, [95 % CI, 0.53, 0.97] and 0.87 [95 % CI, 0.81, 0.94], respectively). OHCA patients receiving manual resuscitation were more likely to attain ROSC compared with load-distributing bands chest compression device (RR 0.88, [95 % CI, 0.80, 0.96]). The in-hospital studies suggested increased relative harm with mechanical compressions for ratio of survival to hospital discharge (RR 0.54, [95 % CI 0.29, 0.98]). However, the results were not statistically significant between different kinds of mechanical chest compression devices and manual resuscitation in survival to admission, discharge and CPC scores for OHCA patients and survival to discharge for in-hospital CA patients.
Conclusions
The ability to achieve ROSC with mechanical devise was inferior to manual chest compression during resuscitation. The use of mechanical chest compression cannot be recommended as a replacement for manual CPR, but rather a supplemental treatment in an overall strategy for treating CA patients.
【 授权许可】
2016 Li et al.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
20160218084709407.pdf | 1333KB | download | |
Fig. 9. | 17KB | Image | download |
Fig. 8. | 17KB | Image | download |
Fig. 7. | 29KB | Image | download |
Fig. 6. | 22KB | Image | download |
Fig. 5. | 31KB | Image | download |
Fig. 4. | 29KB | Image | download |
Fig. 3. | 22KB | Image | download |
Fig. 2. | 24KB | Image | download |
Fig. 1. | 31KB | Image | download |
【 图 表 】
Fig. 1.
Fig. 2.
Fig. 3.
Fig. 4.
Fig. 5.
Fig. 6.
Fig. 7.
Fig. 8.
Fig. 9.
【 参考文献 】
- [1]Ewy GA. Cardiocerebral resuscitation: the new cardiopulmonary resuscitation. Circulation. 2005; 111(16):2134-42.
- [2]Kern KB, Carter AB, Showen RL, Voorhees WD, Babbs CF, Tacker WA et al.. Twenty-four hour survival in a canine model of cardiac arrest comparing three methods of manual cardiopulmonary resuscitation. J Am Coll Cardiol. 1986; 7:859-67.
- [3]Maier GW, Tyson GS, Olsen CO, Kernstein KH, Davis JW, Conn EH et al.. The physiology of external cardiac massage: high-impulse cardiopulmonary resuscitation. Circulation. 1984; 70:86-101.
- [4]Swart GL, Mateer JR, DeBehnke DJ, Jameson SJ, Osborn JL. The effect of compression duration on hemodynamics during mechanical high-impulse CPR. Acad Emerg Med. 1994; 1:430-7.
- [5]Havel C, Schreiber W, Riedmuller E, Haugk M, Richling N, Trimmel H et al.. Quality of closed chest compression in ambulance vehicles, flying helicopters and at the scene. Resuscitation. 2007; 73:264-70.
- [6]Odegaard S, Olasveengen T, Steen PA, Kramer-Johansen J. The effect of transport on quality of cardiopulmonary resuscitation in out-of-hospital cardiac arrest. Resuscitation. 2009; 80:843-8.
- [7]Olasveengen TM, Wik L, Steen PA. Quality of cardiopulmonary resuscitation before and during transport in out-of-hospital cardiac arrest. Resuscitation. 2008; 76:185-90.
- [8]Foo NP, Chang JH, Su SB, Lin HJ, Chen KT, Cheng CF et al.. A stabilization device to improve the quality of cardiopulmonary resuscitation during ambulance transportation: a randomized crossover trial. Resuscitation. 2013; 84:1579-84.
- [9]Abella BS, Alvarado JP, Myklebust H, Edelson DP, Barry A, O’Hearn N et al.. Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest. J Am Med Assoc. 2005; 293(3):305-10.
- [10]Abella BS, Sandbo N, Vassilatos P, Alvarado JP, O’Hearn N, Wigder HN et al.. Chest compression rates during cardiopulmonary resuscitation are suboptimal: a prospective study during in-hospital cardiac arrest. Circulation. 2005; 111(4):428-34.
- [11]Ko PC, Chen WJ, Lin CH, Ma MH, Lin FY. Evaluating the quality of prehospital cardiopulmonary resuscitation by reviewing automated external defibrillator records and survival for out-of-hospital witnessed arrests. Resuscitation. 2005; 64(2):163-9.
- [12]Wik L, Kramer-Johansen J, Myklebust H, Sørebø H, Svensson L, Fellows B et al.. Quality of cardiopulmonary resuscitation during out-of-hospital cardiac arrest. J Am Med Assoc. 2005; 293(3):299-304.
- [13]Lurie K. Mechanical devices for cardiopulmonary resuscitation: an update. Emerg Med Clin North Am. 2002; 20(4):771-84.
- [14]Ikeno F, Kaneda H, Hongo Y, Sakanoue Y, Nolasco C, Emami S et al.. Augmentation of tissue perfusion by a novel compression device increases neurological intact survival in porcine model of prolonged cardiac arrest. Resuscitation. 2006; 68:109-18.
- [15]Halperin H, Paradis N, Ornato J. Cardiopulmonary resuscitation with a novel chest compression device in a porcine model of cardiac arrest: improved hemodynamics and mechanisms. J Am Coll Cardiol. 2004; 44:2214-20.
- [16]Casner M, Andersen D, Isaacs SM. The impact of a new CPR assist device on rate of return of spontaneous circulation in out-of-hospital cardiac arrest. Prehosp Emerg Care. 2005; 9:61-7.
- [17]Lewis RJ, Niemann JT. Manual vs device-assisted CPR: reconciling apparently contradictory results. J Am Med Assoc. 2006; 295(22):2661-4.
- [18]Swanson M, Poniatowski M, O’Keefe M, Shedd JG. Effect of a CPR assist device on survival to emergency department arrival in out of hospital cardiac arrest. Circulation. 2005; 112(17):U1186.
- [19]Swanson M, Poniatowski M, O’Keefe M, Springer P. A CPR assist device increased emergency department admission and end tidal carbon dioxide partial pressures during treatment of out of hospital cardiac arrest. Circulation. 2006; 114(18):554.
- [20]Westfall M, Krantz S, Mullin C, Kaufman C. Mechanical versus manual chest compressions in out-of-hospital cardiac arrest: a meta-analysis. Crit Care Med. 2013; 41(7):1782-9.
- [21]Ong ME, Mackey KE, Zhang ZC, Tanaka H, Ma MH, Swor R. Mechanical CPR devices compared to manual CPR during out-of-hospital cardiac arrest and ambulance transport: a systematic review. Scand J Trauma Resusc Emerg Med. 2012; 20:39. BioMed Central Full Text
- [22]Brooks SC, Hassan N, Bigham BL, Morrison LJ. Mechanical versus manual chest compressions for cardiac arrest. Cochrane Database Syst Rev. 2014; 2:CD007260.
- [23]Dickinson ET, Verdile VP, Schneider RM, Salluzzo RF. Effectiveness of mechanical versus manual chest compressions in out-of-hospital cardiac arrest resuscitation: a pilot study. Am J Emerg Med. 1998; 16(3):289-92.
- [24]Halperin HR, Tsitlik JE, Gelfand M, Weisfeldt ML, Gruben KG, Levin HR. A preliminary study of cardiopulmonary resuscitation by circumferential compression of the chest with use of a pneumatic vest. N Engl J Med. 1993; 329(11):762-8.
- [25]Taylor GJ, Rubin R, Tucker M, Greene HL, Rudikoff MT, Weisfeldt ML. External cardiac compression: a randomized comparison of mechanical and manual techniques. J Am Med Assoc. 1978; 240(7):644-6.
- [26]Hallstrom A, Rea TD, Sayre MR, Christenson J, Anton AR, Mosesso VN et al.. Manual chest compression vs use of an automated chest compression device during resuscitation following out-of-hospital cardiac arrest: a randomized trial. J Am Med Assoc. 2006; H295(22):2620-8.
- [27]Lu XG, Kang X, Gong DB. The clinical efficacy of Thumper modal 1007 cardiopulmonary resuscitation: a prospective randomized control trial. Zhongguo Wei Zhong Bing Ji Jiu Yi Xue. 2010; 22(8):496-7.
- [28]Smekal D, Johansson J, Huzevka T, Rubertsson S. A pilot study of mechanical chest compressions with the LUCASTM device in cardiopulmonary resuscitation. Resuscitation. 2011; 82(6):702-6.
- [29]Axelsson C, Nestin J, Svensson L, Axelsson AB, Herlitz J. Clinical consequences of the introduction of mechanical chest compression in the EMS system for treatment of out-of-hospital cardiac arrest-a pilot study. Resuscitation. 2006; 71(1):47-55.
- [30]Hock Ong ME, Fook-Chong S, Annathurai A, Ang SH, Tiah L, Yong KL. Improved neurologically intact survival with the use of an automated, load-distributing band chest compression device for cardiac arrest presenting to the emergency department. Crit Care. 2012; 16:R144. BioMed Central Full Text
- [31]Ong ME, Ornato JP, Edwards DP, Dhindsa HS, Best AM, Ines CS. Use of an automated, load-distributing band chest compression device for out-of-hospital cardiac arrest resuscitation. J Am Med Assoc. 2006; 295(22):2629-37.
- [32]Zeiner S, Sulzgruber P, Datler P, Keferböck M, Poppe M, Lobmeyr E et al.. Chest compression does not seem to improve outcome after out-of hospital cardiac arrest. A single center observational trial. Resuscitation. 2015; 96:220-5.
- [33]Rubertsson S, Lindgren E, Smekal D, Ostlund O, Silfverstolpe J, Robert A et al.. Mechanical chest compressions and simultaneous defibrillation vs conventional cardiopulmonary resuscitation in out-of-hospital cardiac arrest. J Am Med Assoc. 2014; 311(1):53-61.
- [34]Perkins GD, Lall R, Quinn T, Deakin CD, Cooke MW, Horton J et al.. Mechanical versus manual chest compression for out-of-hospital cardiac arrest (PARAMEDIC): a pragmatic, cluster randomised controlled trial. Lancet. 2015; 385:947-55.
- [35]Kern KB, Hilwig RW, Berg RA, Sanders AB, Ewy GA. Importance of continuous chest compressions during cardiopulmonary resuscitation: improved outcome during a simulated single lay-rescuer scenario. Circulation. 2002; 105(5):645-9.
- [36]Yu T, Weil MH, Tang W, Sun S, Klouche K, Povoas H et al.. Adverse outcomes of interrupted precordial compression during automated defibrillation. Circulation. 2002; 106(3):368-72.
- [37]Eftestol T, Sunde K, Steen PA. Effects of interrupting precordial compressions on the calculated probability of defibrillation success during out-of-hospital cardiac arrest. Circulation. 2002; 105(19):2270-3.
- [38]Berg MD, Schexnayder SM, Chameides L, Terry M, Donoghue A, Hickey RW et al.. Part 5: Adult Basic Life Support: 2010 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2010; 122 Suppl 3:S685-705.
- [39]Hightower D, Thomas SH, Stone CK, Dunn K, March JA. Decay in quality of closed-chest compressions over time. Ann Emerg Med. 1995; 26(3):300-3.
- [40]Gatesa S, Quinnb T, Deakinc CD, Blaire L, Coupera K, Perkins GD. Mechanical chest compression for out of hospital cardiac arrest: systematic review and meta-analysis. Resuscitation. 2015; 94:91-7.
- [41]Herlitz J, Ekstrom L, Wennerblom B, Axelsson A, Bang A, Holmberg S. Effect of bystander initiated cardiopulmonary resuscitation on ventricular fibrillation and survival after witnessed cardiac arrest outside hospital. Br Heart J. 1994; 72(5):408-12.
- [42]McCarney R, Warner J, Iliffe S, van Haselen R, Griffin M, Fisher P. The Hawthorne Effect: a randomised, controlled trial. BMC Med Res Methodol. 2007; 7:30. BioMed Central Full Text
- [43]Fox NS, Brennan JS, Chasen ST. Clinical estimation of fetal weight and the Hawthorne effect. Eur J Obstet Gynecol Reprod Biol. 2008; 141(2):111-4.
- [44]Valenzuela TD, Roe DJ, Cretin S, Spaite DW, Larsen MP. Estimating effectiveness of cardiac arrest interventions: a logistic regression survival model. Circulation. 1997; 96(10):3308-13.
- [45]Wik L, Hansen TB, Fylling F, Steen T, Vaagenes P, Auestad BH et al.. Delaying defibrillation to give basic cardiopulmonary resuscitation to patients with out-of-hospital ventricular fibrillation: a randomized trial. J Am Med Assoc. 2003; 289:1389-95.