Journal of Otolaryngology-Head & Neck Surgery | |
Preliminary report: neural firing patterns specific for Meniere’s disease | |
Zahra Moussavi4  Brian Lithgow1  Zeinab A Dastgheib3  Brian Blakley2  | |
[1] Monash Alfred Psychiatry Research Centre, University of Manitoba, and Research Affiliate of Riverview Health Center, EVestG Research Lab, Riverview Health Centre, Room PE446, 1 Morley Avenue, Winnipeg R3L2P4, MB, Canada;Department of Otolaryngology - Head and Neck Surgery, University of Manitoba, GB421 - 820 Sherbrook Street, Winnipeg R3A 1R9, Manitoba, Canada;University of Manitoba, Room E3-512 Eng. Bldg., Biomedical Engineering, 75A Chancellor’s Circle, Winnipeg R3T 5 V6, MB, Canada;Biomedical Engineering, University of Manitoba, and Research Affiliate of Riverview Health Center, Room E3-513 Eng. Bldg., Biomedical Engineering, 75A Chancellor’s Circle, Winnipeg R3T 5V6, MB, Canada | |
关键词: Fractal dimension; Classification; Vestibular response; EVestG; Meniere’s disease; | |
Others : 1132942 DOI : 10.1186/s40463-014-0052-4 |
|
received in 2014-06-18, accepted in 2014-12-08, 发布年份 2014 | |
【 摘 要 】
Objective
To describe the application of a new, objective diagnostic test for Meniere’s disease.
Introduction
Electrovestibulography (EVestG) is a complex, newly-developed test paradigm that searches for neural firing patterns that may be diagnostic for particular neural disorders. EVestG system was previously “trained” to distinguish Meniere’s disease from other patients on a set of training data. In this paper we illustrate its diagnostic application in a new group of unknown subjects.
Setting
Collaborative Academic Bioengineering Research Centre.
Study design
Prospective, blinded human Clinical Trial.
Methods
In an attempt to understand the specific neural firing patterns that may objectively characterize latent Meniere’s disease, two hundred fifty-six consecutive patients who presented for electronystagmography testing were asked to undergo EVestG testing. Ten subjects actually completed testing but data were too noisy to permit analysis for one patient. Complete data were available for nine patients with either a clinical diagnosis of either Meniere’s disease (4 patients) or some other vestibular disorder (2 vestibular neuritis, 2 benign positional vertigo and 1 non-specific dizziness). None of the patients were experiencing attacks of vertigo within a week of EVestG testing. Ten normal control subjects with no history or symptoms of ear disease were also tested. EVestG was performed in a separate engineering research facility by investigators who were unaware of their clinical diagnosis. If EVestG suggested that the probability of Meniere’s disease was 0.5 or greater Meniere’s disease was considered present by the objective testing. The objective and clinical diagnoses were compared.
Results
EVestG testing correctly identified three of four Meniere’s disease patients and rejected the diagnosis in 9 of the 10 controls. Two of the 5 dizzy, non-Meniere’s patients were incorrectly identified as Meniere’s disease. The sensitivity and specificity of EvestG testing were 75% and 80%, respectively. EVestG results were statistically significantly different for Meniere’s patients versus the other dizzy patients and controls (Univariate ANOVA difference contrasts p = 0.0340) even in this small sample.
Conclusion
The EVestG protocol appeared to show promise as an objective, diagnostic test for Meniere’s disease, but our sample size is too small to generalize widely.
Level of evidence
N.A. Prospective Human clinical trial.
【 授权许可】
2014 Blakley et al.; licensee BioMed Central.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
20150304095942991.pdf | 393KB | download | |
Figure 1. | 26KB | Image | download |
【 图 表 】
Figure 1.
【 参考文献 】
- [1]Baloh RW, Kerber KA: Clinical Neurophysiology of the Vestibular System. Oxford University Press, New York; 2011.
- [2]Committee on Hearing and Equilibrium guidelines for the diagnosis and evaluation of therapy in Meniere's disease: American Academy of Otolaryngology-Head and Neck Foundation, Inc Otolaryngol Head Neck Surg 1995, 113:181-185.
- [3]Lithgow B: A Methodology for Detecting Field Potentials from the External Ear Canal: NEER and EVestG. Ann Biomed Eng 2012, 40(8):1835-1850. 2012.:1835–1850
- [4]Lithgow BJ, Garrett A, Heibert D: EVestG: a measure for Meniere’s Disease. Conf Proc IEEE Eng Med Biol Soc 2008, 2008:4162-4165.
- [5]Dastgheib Z, Lithgow B, Blakley B, Moussavi Z. A New Diagnostic Vestibular Evoked Response Journal of Otolaryngology-Head and Neck Surgery 2014; In Press Oct. 22, 2014.
- [6]Hogg R, Ledolter J: Engineering statistics. MacMillan, New York; 1987.
- [7]Leech NL, Barrett KC, GA M: SPSS for Intermediate Statistics – Use and Interpretation. Psychology Press, New York NY; 2008.