Fully integrated, fully automated generation of short tandem repeat profiles
Richard F Selden1  James W Schumm1  Luke Palombo1  Sameer Vasantgadkar1  Catherine Hogan1  Rosemary S Turingan1  Eugene Tan1 
[1] NetBio, 830 Winter Street, Waltham, MA 02451, USA
关键词: Short tandem repeat;    Rapid DNA analysis;    Polymerase chain reaction;    Microfluidics;    CODIS;    Automation;   
Others  :  801469
DOI  :  10.1186/2041-2223-4-16
 received in 2013-05-25, accepted in 2013-07-18,  发布年份 2013
PDF
【 摘 要 】

Background

The generation of short tandem repeat profiles, also referred to as ‘DNA typing,’ is not currently performed outside the laboratory because the process requires highly skilled technical operators and a controlled laboratory environment and infrastructure with several specialized instruments. The goal of this work was to develop a fully integrated system for the automated generation of short tandem repeat profiles from buccal swab samples, to improve forensic laboratory process flow as well as to enable short tandem repeat profile generation to be performed in police stations and in field-forward military, intelligence, and homeland security settings.

Results

An integrated system was developed consisting of an injection-molded microfluidic BioChipSet cassette, a ruggedized instrument, and expert system software. For each of five buccal swabs, the system purifies DNA using guanidinium-based lysis and silica binding, amplifies 15 short tandem repeat loci and the amelogenin locus, electrophoretically separates the resulting amplicons, and generates a profile. No operator processing of the samples is required, and the time from swab insertion to profile generation is 84 minutes. All required reagents are contained within the BioChipSet cassette; these consist of a lyophilized polymerase chain reaction mix and liquids for purification and electrophoretic separation.

Profiles obtained from fully automated runs demonstrate that the integrated system generates concordant short tandem repeat profiles. The system exhibits single-base resolution from 100 to greater than 500 bases, with inter-run precision with a standard deviation of ±0.05 - 0.10 bases for most alleles. The reagents are stable for at least 6 months at 22°C, and the instrument has been designed and tested to Military Standard 810F for shock and vibration ruggedization. A nontechnical user can operate the system within or outside the laboratory.

Conclusions

The integrated system represents the first generation of a turnkey approach to short tandem repeat profiling and has the potential for use in both the field (for example, police booking stations, the battlefield, borders and ports) and the forensic laboratory.

【 授权许可】

   
2013 Tan et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20140708010100317.pdf 3569KB PDF download
Figure 11. 88KB Image download
Figure 10. 133KB Image download
Figure 9. 30KB Image download
Figure 8. 244KB Image download
Figure 7. 45KB Image download
Figure 6. 34KB Image download
Figure 5. 33KB Image download
Figure 4. 78KB Image download
Figure 3. 56KB Image download
Figure 2. 33KB Image download
Figure 1. 67KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

Figure 4.

Figure 5.

Figure 6.

Figure 7.

Figure 8.

Figure 9.

Figure 10.

Figure 11.

【 参考文献 】
  • [1]Jeffreys AJ, Wilson V, Thein SL: Hypervariable ‘minisatellite’ regions in human DNA. Nature 1985, 314:67-73.
  • [2]Jeffreys AJ, Wilson V, Thein SL: Individual-specific ‘fingerprints’ of human DNA. Nature 1985, 316:76-79.
  • [3]Jeffreys AJ, Brookfield JFY, Semeonoff R: Positive identification of an immigration test-case using human DNA fingerprints. Nature 1985, 317:818-819.
  • [4]Weber JL, May PE: Abundant class of human DNA polymorphisms which can be typed using the polymerase chain reaction. Am J Hum Genet 1989, 44:388.
  • [5]Litt M, Luty JA: A hypervariable microsatellite revealed by in vitro amplification of a dinucleotide repeat within the cardiac muscle actin gene. Am J Hum Genet 1989, 44:397.
  • [6]Rassmann K, Schlotterer C, Tautz D: Isolation of simple-sequence loci for use in polymerase chain reaction-based DNA fingerprinting. Electrophoresis 1991, 12:113-118.
  • [7]Edwards A, Civitello A, Hammond HA, Caskey CT: DNA typing and genetic mapping with trimeric and tetrameric tandem repeats. Am J Hum Genet 1991, 49:746-756.
  • [8]Budowle B: Studies for selecting core STR loci for CODIS. In In Proceedings of Cambridge Healthtech Institute’s Second Annual Conference on DNA Forensics: Science, Evidence and Future Prospects.. McLean, VA:: Cambridge Health Tech Institute; 1997;
  • [9]Moretti T, Budowle B: The CODIS STR project: evaluation of fluorescent multiplex STR systems. In Proceedings of the 50th Annual Meeting of the American Academy of Forensic Sciences. San Francisco, CA: American Academy of Forensic Sciences; 1998.
  • [10]U.S. Department of Defense Biometric and Forensic Technology Forum. [http://csis.org/event/us-department-defense-biometric-and-forensic-technology-forum webcite]
  • [11]Bienvenue JM, Legendre LA, Ferrance JP, Landers JP: An integrated microfluidic device for DNA purification and PCR amplification of STR fragments. Forensic Sci Int Genet 2010, 4:178-186.
  • [12]El-Sissi O, Franklin H, Nielsen B, Pagano S, Belcinski R, Bogdan G, Jovanuvich S: Apollo 200 fully integrated DNA HID system: multi channel results in under 2 hours. J Biomol Tech 2011, 22:S44.
  • [13]Hopwood AJ, Hurth C, Yang J, Cai Z, Moran N, Lee-Edghill JG, Nordquist A, Lenigk R, Estes MD, Haley JP: Integrated microfluidic system for rapid forensic DNA analysis: sample collection to DNA profile. Anal Chem 2010, 82:6991-6999.
  • [14]Boom R, Sol CJ, Salimans MM, Jansen CL, Wertheim-van Dillen PM, van der-Noordaa J: Rapid and simple method for purification of nucleic acids. J Clin Microbiol 1990, 28:495-503.
  • [15]Giese H, Lam R, Selden R, Tan E: Fast multiplexed polymerase chain reaction for conventional and microfluidic short tandem repeat analysis. J Forensic Sci 2009, 54:1287-1296.
  • [16]Bar W, Brinkmann B, Budowle B, Carracedo A, Gill P, Lincoln P, Mayr W, Olaisen B: DNA recommendations. Further report of the DNA Commission of the ISFH regarding the use of short tandem repeat systems. International Society for Forensic Haemogenetics. Int J Legal Med 1997, 110:175-176.
  • [17]Lincoln P: DNA recommendations: further report of the DNA Commission of the ISFH regarding the use of short tandem repeat systems. Forensic Sci Int 1997, 87:179-184.
  • [18]Luckey JA, Norris TB, Smith LM: Analysis of resolution in DNA sequencing by capillary gel electrophoresis. J Phys Chem 1993, 97:3067-3075.
  • [19]Schumm J, Gutierrez-Mateo C, Tan E, Selden RF: A 27-locus STR assay to meet all United States and European law enforcement agency standards. J Forensic Sci 2013.
  • [20]Hares DR: Expanding the CODIS core loci in the United States. Forensic Sci Int Genet 2012, 6:e52-e54.
  • [21]Hares DR: Addendum to expanding the CODIS core loci in the United States. Forensic Sci Int Genet 2012, 6:e135.
  • [22]Huang S, Zhu Y, Shen X, Le X, Yan H: Genetic variation analysis of 15 autosomal STR loci of AmpFlSTR Sinofiler (TM) PCR amplification kit in Henan (central China) Han population. Leg Med 2010, 12:160-161.
  • [23]Jacewicz R, Jedrzejczyk M, Ludwikowska M, Berent J: Polymorphism of pentanucleotide STR markers in Polish population sample. Forensic Sci Int: Gen Suppl Series 2008, 1:334-336.
  • [24]Scientific Working Group on DNA Analysis Methods (SWGDAM): Validation Guidelines for DNA Analysis Methods. [http://swgdam.org/SWGDAM_Validation_Guidelines_APPROVED_Dec_2012.pdf webcite]
  • [25]Supreme Court of the United States: Maryland v King 569 U.S. No. 12–207 (2013). [http://www.supremecourt.gov/opinions/12pdf/12-207_d18e.pdf webcite]
  文献评价指标  
  下载次数:77次 浏览次数:48次