期刊论文详细信息
BMC Medical Imaging
Correlation of clinical and physical-technical image quality in chest CT: a human cadaver study applied on iterative reconstruction
Klaus Bacher1  Hubert Thierens1  Katharina D’Herde1  Koenraad Verstraete2  Eric Achten2  Mathias Van Borsel2  Tom Dewaele2  Merel Vergauwen2  Tom Van Hoof1  Peter Smeets2  An De Crop1 
[1] Department of Basic Medical Sciences, Ghent University, Proeftuinstraat 86, Ghent, B-9000, Belgium;Department of Radiology, Ghent University Hospital, De Pintelaan 185, Ghent, B-9000, Belgium
关键词: Visual grading analysis;    Human cadaver study;    Iterative reconstruction;    Image quality;    Chest CT;   
Others  :  1222828
DOI  :  10.1186/s12880-015-0075-y
 received in 2014-12-02, accepted in 2015-08-10,  发布年份 2015
PDF
【 摘 要 】

Background

The first aim of this study was to evaluate the correlation between clinical and physical-technical image quality applied to different strengths of iterative reconstruction in chest CT images using Thiel cadaver acquisitions and Catphan images. The second aim was to determine the potential dose reduction of iterative reconstruction compared to conventional filtered back projection based on different clinical and physical-technical image quality parameters.

Methods

Clinical image quality was assessed using three Thiel embalmed human cadavers. A Catphan phantom was used to assess physical-technical image quality parameters such as noise, contrast-detail and contrast-to-noise ratio (CNR).

Both Catphan and chest Thiel CT images were acquired on a multislice CT scanner at 120 kVp and 0.9 pitch. Six different refmAs settings were applied (12, 30, 60, 90, 120 and 150refmAs) and each scan was reconstructed using filtered back projection (FBP) and iterative reconstruction (SAFIRE) algorithms (1,3 and 5 strengths) using a sharp kernel, resulting in 24 image series. Four radiologists assessed the clinical image quality, using a visual grading analysis (VGA) technique based on the European Quality Criteria for Chest CT.

Results

Correlation coefficients between clinical and physical-technical image quality varied from 0.88 to 0.92, depending on the selected physical-technical parameter. Depending on the strength of SAFIRE, the potential dose reduction based on noise, CNR and the inverse image quality figure (IQF inv)varied from 14.0 to 67.8 %, 16.0 to 71.5 % and 22.7 to 50.6 % respectively. Potential dose reduction based on clinical image quality varied from 27 to 37.4 %, depending on the strength of SAFIRE.

Conclusion

Our results demonstrate that noise assessments in a uniform phantom overestimate the potential dose reduction for the SAFIRE IR algorithm. Since the IQF invbased dose reduction is quite consistent with the clinical based dose reduction, an optimised contrast-detail phantom could improve the use of contrast-detail analysis for image quality assessment in chest CT imaging. In conclusion, one should be cautious to evaluate the performance of CT equipment taking into account only physical-technical parameters as noise and CNR, as this might give an incomplete representation of the actual clinical image quality performance.

【 授权许可】

   
2015 De Crop et al.

【 预 览 】
附件列表
Files Size Format View
20150828021248665.pdf 1367KB PDF download
20141203093118369.pdf 1261KB PDF download
Fig. 6. 20KB Image download
Fig. 5. 18KB Image download
Fig. 4. 19KB Image download
Fig. 3. 33KB Image download
Fig. 2. 18KB Image download
Fig. 1. 39KB Image download
【 图 表 】

Fig. 1.

Fig. 2.

Fig. 3.

Fig. 4.

Fig. 5.

Fig. 6.

【 参考文献 】
  • [1]Publication 87: Managing patient dose in computed tomography. 2000.
  • [2]Brenner DJ, Hall EJ. Computed tomography - an increasing source of radiation exposure. N Engl J Med. 2007; 357:2277-2284.
  • [3]Deak PD et al.. Effects of adaptive section collimation on patient radiation dose in multisection spiral CT. Radiology. 2009; 252(1):140-147.
  • [4]Lee T-Y, Chhem RK. Impact of new technologies on dose reduction in CT. Eur J Radiol. 2010; 76:28-35.
  • [5]Kalra MK et al.. Strategies for CT radiation dose optimization. Radiology. 2004; 230(3):619-28.
  • [6]Kubo T et al.. Radiation dose reduction in chest CT: a review. AJR Am J Roentgenol. 2008; 190(2):335-43.
  • [7]Solomon J, Samei E. Quantum noise properties of CT images with anatomical textured backgrounds across reconstruction algorithms: FBP and SAFIRE. Med Phys. 2014; 41(9):12.
  • [8]Mieville FA et al.. Paediatric cardiac CT examinations: impact of the iterative reconstruction method ASIR on image quality - preliminary findings. Pediatr Radiol. 2011; 41(9):1154-1164.
  • [9]Martinsen AC et al.. Iterative reconstruction reduces abdominal CT dose. Eur J Radiol. 2012; 81(7):1483-7.
  • [10]Veldkamp W, Kroft L, Geleijns J. Dose and perceived image quality in chest radiography. Eur J Radiol. 2009; 72:209-217.
  • [11]Thiel W. Die Konservierung ganzer Leichen in natürlichen Farben. Ann Anat. 1992; 174:185-195.
  • [12]Thiel W. Ergänzung für die Konservierung ganze Leichen nach W. Thiel. Ann Anat. 2002; 184:267-269.
  • [13]De Crop A et al.. Correlation of contrast-detail analysis and clinical image quality assessment in chest radiography with a human cadaver study. Radiology. 2012; 262(1):298-304.
  • [14]AAPM report no.39, Specification and acceptance testing of computed tomography scanners. AAPM, 1993.
  • [15]Samei E et al.. Assessment of display performance for medical imaging systems: executive summary of AAPM TG18 report. Med Phys. 2005; 32(4):1205-1225.
  • [16]Commission of the European Communities, European guidelines on quality criteria for computed tomography (EUR 16262). 1999.
  • [17]CEC quality criteria for diagnostic radiographic images and patient exposure trial. (EUR 12952 EN). 1990.
  • [18]Sund P et al.. Comparison of visual grading analysis and determination of detective quantum efficiency for evaluation system performance in digital chest radiography. Eur Radiol. 2004; 14:48-58.
  • [19]Borjesson S et al.. A software tool for increased efficiency in observer performance studies in radiology. Radiat Prot Dosim. 2005; 114(1–3):45-52.
  • [20]Thijssen M, Bijkerk K, van der Burgth R. Manual Contrast-Detail Phantom CDRAD type 2.0. Project Quality Assurance in Radiology, Department of Radiology, University Hospital Nijmegen, St. Radboud, The Netherlands. 1998.
  • [21]Viner M et al.. Liver SULmean at FDG PET/CT: Interreader Agreement and Impact of Placement of Volume of Interest. Radiology. 2013; 267(2):596-601.
  • [22]Bath M. Evaluating imaging systems: practical applications. Radiat Prot Dosimetry. 2010; 139(1–3):26-36.
  • [23]Baker ME et al.. Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver. Am J Roentgenol. 2012; 199(1):8-18.
  • [24]Barrett HH et al.. Model observers for assessment of image quality. Proc Natl Acad Sci U S A. 1993; 90(21):9758-9765.
  • [25]Richard S, Siewerdsen JH. Comparison of model and human observer performance for detection and discrimination tasks using dual-energy x-ray images. Med Phys. 2008; 35(11):5043-5053.
  • [26]Zhang Y et al.. Correlation between human and model observer performance for discrimination task in CT. Phys Med Biol. 2014; 59(13):3389-3404.
  • [27]Yip M et al.. Validation of a simulated dose reduction methodology using digital mammography CDMAM images and mastectomy images. Digital Mammography. 2010; 6136:78-85.
  • [28]Veldkamp W et al.. Contrast-detail evaluation and dose assessment of eight digital chest radiography systems in clinical practice. Eur Radiol. 2006; 16:333-341.
  • [29]Mieville FA et al.. Iterative reconstruction methods in two different MDCT scanners: physical metrics and 4-alternative forced-choice detectability experiments - a phantom approach. Phys Med. 2013; 29(1):99-110.
  • [30]Kim M et al.. Adaptive iterative dose reduction algorithm in CT: effect on image quality compared with filtered back projection in body phantoms of different sizes. Korean J Radiol. 2014; 15(2):195-204.
  • [31]Ghetti C, Ortenzia O, Serreli G. CT iterative reconstruction in image space: a phantom study. Phys Med. 2012; 28(2):161-165.
  • [32]Ghetti C et al.. Physical characterization of a new CT iterative reconstruction method operating in sinogram space. J Appl Clin Med Phys. 2013; 14(4):263-271.
  • [33]Li K et al.. Statistical model based iterative reconstruction (MBIR) in clinical CT systems. Part II. Experimental assessment of spatial resolution performance. Med Phys. 2014; 41(7):12.
  • [34]Prakash P et al.. Radiation dose reduction with chest computed tomography using adaptive statistical iterative reconstruction technique: initial experience. J Comput Assist Tomogr. 2010; 34(1):40-45.
  • [35]Pontana F et al.. Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2): image quality of low-dose CT examinations in 80 patients. Eur Radiol. 2011; 21(3):636-643.
  文献评价指标  
  下载次数:43次 浏览次数:8次