Quality assessment of digital X-ray chest images using an anthropomorphic chest phantom | |
Vodovatov, A.V.^1 ; Kamishanskaya, I.G.^2 ; Drozdov, A.A.^2 ; Bernhardsson, C.^3 | |
Saint Petersburg Research Institute of Radiation Hygiene, Saint-Petersburg | |
197101, Russia^1 | |
Saint Petersburg State University, Faculty of Medicine, Saint-Petersburg, Russia^2 | |
Medical Radiation Physics, Department of Translational Medicine, Malmö, Lund University, Skåne University Hospital, Malmö | |
205 02, Sweden^3 | |
关键词: Anthropomorphic phantoms; Chest phantom; Human anatomy; Quality assessment; Screening examinations; Tube voltages; Visual grading analysis; X-ray image; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/808/1/012009/pdf DOI : 10.1088/1742-6596/808/1/012009 |
|
来源: IOP | |
【 摘 要 】
The current study is focused on determining the optimal tube voltage for the conventional X-ray digital chest screening examinations, using a visual grading analysis method. Chest images of an anthropomorphic phantom were acquired in posterior-anterior projection on four digital X-ray units with different detector types. X-ray images obtained with an anthropomorphic phantom were accepted by the radiologists as corresponding to a normal human anatomy, hence allowing using phantoms in image quality trials without limitations.