Leaks can dramatically decrease FiO2 on home ventilators: a bench study
Eric Meaudre1  Bruno Palmier1  Julien Bordes1  Bertrand Prunet1  Pierre Esnault1  Yves Asencio1  Erwan Daranda1  Philippe Goutorbe1 
[1] Emergency, Anesthesia and ICU Departement, Military Teaching Hospital Sainte Anne, Boulevard Sainte Anne, Toulon 83000, France
关键词: FiO2;    COPD;    Home ventilation;    Leaks;    NIV;   
Others  :  1142118
DOI  :  10.1186/1756-0500-6-282
 received in 2013-05-06, accepted in 2013-07-18,  发布年份 2013
PDF
【 摘 要 】

Background

Long term oxygen therapy improves survival in hypoxemic patients with chronic obstructive pulmonary disease (COPD). Because pressure support ventilation with a home care ventilator is largely unsupervised, there is considerable risk of leakage occurring, which could affect delivered FiO2. We have therefore conducted a bench study in order to measure the effect of different levels of O2 supply and degrees of leakage on delivered FiO2. Ventilator tested: Legendair® (Airox™, Pau, France). Thirty-six measures were performed in each four ventilators with zero, 5 and 10 l.min-1 leakage and 1,2,4 and 8 l O2 flow.

Findings

FiO2 decreased significantly with 5 l.min-1 leakage for all O2 flow rates, and with 10 l.min-1 at 4 and 8 l.min-1 O2.

Conclusion

During application of NIV on home ventilators, leakage can dramatically decrease inspired FiO2 making it less effective. It is important to know the FiO2 dispensed when NIV is used for COPD at home. We would encourage industry to develop methods for FiO2 regulation Chronic use of NIV for COPD with controlled FiO2 or SpO2 requires further studys.

【 授权许可】

   
2013 Goutorbe et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150327222723990.pdf 487KB PDF download
Figure 2. 54KB Image download
Figure 1. 69KB Image download
【 图 表 】

Figure 1.

Figure 2.

【 参考文献 】
  • [1]Continuous or Nocturnal Oxygen Therapy in Hypoxemic Chronic Obstructive Lung Disease [Internet] annals.org. 2012. 2012 [cited 2012 Feb 8]. Available from: http://www.annals.org/content/93/3/391.abstract?sid=79927bd3-e7f2-4273-baab-19b98e3d4991 webcite
  • [2]Council M: Long term domiciliary oxygen therapy in chronic hypoxic cor pulmonale complicating chronic bronchitis and emphysema: Report of the Medical Research Council Working Party. Lancet 1981, 1(8222):681-686.
  • [3]Rossi A: Noninvasive ventilation has not been shown to be ineffective in stable COPD. Am J Respir Crit Care Med 2000, 161(3 Pt 1):688-689.
  • [4]Hill NS: Noninvasive ventilation has been shown to be ineffective in stable COPD. Am J Respir Crit Care Med 2000, 161(3):689-691.
  • [5]Casanova C, Celli BR, Tost L, Soriano E, Abreu J, Velasco V, et al.: Long-term controlled trial of nocturnal nasal positive pressure ventilation in patients with severe COPD. Chest 2000, 118(6):1582-1590.
  • [6]Clini E, Sturani C, Rossi A, Viaggi S, Corrado A, Donner CF, Ambrosino N, Rehabilitation and Chronic Care Study Group: The italian multicentre study on noninvasive ventilation in chronic obstructive pulmonary diseases patients. Eur Respir J 2002, 20(3):529-38. Erratum in: Eur Respir J. 2002 Dec;20(6):1617
  • [7]Cuvelier A, Molano LC, Muir JF: Domiciliary ventilation in patients with COPD. Rev Mal Respir 2005, 22(4):615-33.
  • [8]Lloyd-Owen SJ, Donaldson GC, Ambrosino N, Escarabill J, Farre R, Fauroux B, et al.: Patterns of home mechanical ventilation use in Europe: results from the Eurovent survey. Eur Respir J 2005, 25(6):1025-1031.
  • [9]Tassaux D, Strasser S, Fonseca S, Dalmas E, Jolliet P: Comparative bench study of triggering, pressurization, and cycling between the home ventilator VPAP II and three ICU ventilators. Intensive Care Med 2002, 28(9):1254-1261.
  • [10]Richard J-C, Carlucci A, Breton L, Langlais N, Jaber S, Maggiore S, et al.: Bench testing of pressure support ventilation with three different generations of ventilators. Intensive Care Med 2002, 28(8):1049-1057.
  • [11]Kolodziej MA, Jensen L, Rowe B, Sin D: Systematic review of noninvasive positive pressure ventilation in severe stable COPD. Eur Respir J 2007, 30(2):293-306.
  • [12]Diaz O, Begin P, Torrealba B, Jover E, Lisboa C: Effects of noninvasive ventilation on lung hyperinflation in stable hypercapnic COPD. Eur Respir J Eur Respiratory Soc 2002, 20(6):1490-1498.
  • [13]Robert D, Argaud L: Clinical review: long-term noninvasive ventilation. Crit Care 2007, 11(2):210. BioMed Central Full Text
  • [14]Mehta S, Hill NS: Noninvasive ventilation. Am J Respir Crit Care Med Am Thoracic Soc 2001, 163(2):540.
  • [15]Battisti A: Performance Characteristics of 10 Home Mechanical Ventilators in Pressure-Support Mode: A Comparative Bench Study. Chest 2005, 27(5):1784-1792.
  • [16]Schwartz AR, Kacmarek RM, Hess DR: Factors affecting oxygen delivery with bi-level positive airway pressure. Respir Care 2004, 49(3):270-275.
  • [17]Thys F, Liistro G, Dozin O, Marion E, Rodenstein DO: Determinants of Fi, O2 with oxygen supplementation during noninvasive two-level positive pressure ventilation. Eur Respir J 2002, 19(4):653-657.
  • [18]Miyoshi E: Effects of Gas Leak on Triggering Function, Humidification, and Inspiratory Oxygen Fraction During Noninvasive Positive Airway Pressure Ventilation. Chest 2005, 128(5):3691-3698.
  • [19]Chen H, Liang B-M, Xu Z-B, Tang Y-J, Wang K, Xiao J, et al.: Long-term non-invasive positive pressure ventilation in severe stable chronic obstructive pulmonary disease: a meta-analysis. Chin Med J 2011, 124(23):4063-4070.
  文献评价指标  
  下载次数:50次 浏览次数:24次