BioMedical Engineering OnLine | |
Effect of pressure and padding on motion artifact of textile electrodes | |
Alper Cömert1  Markku Honkala2  Jari Hyttinen1  | |
[1] Institute of Biosciences and Medical Technology, Tampere, Finland | |
[2] Department of Material Science, Tampere University of Technology, Tampere, Finland | |
关键词: Electrode-skin interface impedance; Motion artifact; ECG; Textile electrodes; | |
Others : 797914 DOI : 10.1186/1475-925X-12-26 |
|
received in 2012-12-05, accepted in 2013-03-28, 发布年份 2013 | |
【 摘 要 】
Background
With the aging population and rising healthcare costs, wearable monitoring is gaining importance. The motion artifact affecting dry electrodes is one of the main challenges preventing the widespread use of wearable monitoring systems. In this paper we investigate the motion artifact and ways of making a textile electrode more resilient against motion artifact. Our aim is to study the effects of the pressure exerted onto the electrode, and the effects of inserting padding between the applied pressure and the electrode.
Method
We measure real time electrode-skin interface impedance, ECG from two channels, the motion artifact related surface potential, and exerted pressure during controlled motion by a measurement setup designed to estimate the relation of motion artifact to the signals. We use different foam padding materials with various mechanical properties and apply electrode pressures between 5 and 25 mmHg to understand their effect. A QRS and noise detection algorithm based on a modified Pan-Tompkins QRS detection algorithm estimates the electrode behaviour in respect to the motion artifact from two channels; one dominated by the motion artifact and one containing both the motion artifact and the ECG. This procedure enables us to quantify a given setup’s susceptibility to the motion artifact.
Results
Pressure is found to strongly affect signal quality as is the use of padding. In general, the paddings reduce the motion artifact. However the shape and frequency components of the motion artifact vary for different paddings, and their material and physical properties. Electrode impedance at 100 kHz correlates in some cases with the motion artifact but it is not a good predictor of the motion artifact.
Conclusion
From the results of this study, guidelines for improving electrode design regarding padding and pressure can be formulated as paddings are a necessary part of the system for reducing the motion artifact, and further, their effect maximises between 15 mmHg and 20 mmHg of exerted pressure. In addition, we present new methods for evaluating electrode sensitivity to motion, utilizing the detection of noise peaks that fall into the same frequency band as R-peaks.
【 授权许可】
2013 Cömert et al.; licensee BioMed Central Ltd.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
20140706085634348.pdf | 900KB | download | |
Figure 10. | 43KB | Image | download |
Figure 9. | 106KB | Image | download |
Figure 8. | 79KB | Image | download |
Figure 7. | 47KB | Image | download |
Figure 6. | 83KB | Image | download |
Figure 5. | 56KB | Image | download |
Figure 4. | 47KB | Image | download |
Figure 3. | 28KB | Image | download |
Figure 2. | 23KB | Image | download |
Figure 1. | 24KB | Image | download |
【 图 表 】
Figure 1.
Figure 2.
Figure 3.
Figure 4.
Figure 5.
Figure 6.
Figure 7.
Figure 8.
Figure 9.
Figure 10.
【 参考文献 】
- [1]WHO: The top 10 causes of death. http://www.who.int/mediacentre/factsheets/fs310/en/index.html webcite
- [2]Searle A, Kirkup L: A direct comparison of wet, dry and insulating bioelectric recording electrodes. Physiol Meas 2000, 21:271-283.
- [3]Tam H, Webster JG: Minimizing electrode motion artifact by skin abrasion. Biomedical Engineering, IEEE Transactions on 1977, BME-24:134-139.
- [4]Paradiso R, Loriga G, Taccini N: A wearable health care system based on knitted integrated sensors. Information Technology in Biomedicine, IEEE Transactions on 2005, 9:337-344.
- [5]Harris M, Habetha J: The MyHeart project: a framework for personal health care applications. Comput Cardiol 2007, 2007:137-140.
- [6]Noury N, Dittmar A, Corroy C, Baghai R, Weber JL, Blanc D, Klefstat F, Blinovska A, Vaysse S, Comet B: A smart cloth for ambulatory telemonitoring of physiological parameters and activity: the VTAMN project. Enterprise Networking and Computing in Healthcare Industry, 2004. HEALTHCOM 2004, 155-160. [Proceedings. 6th International Workshop on] http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumber=1324507 webcite
- [7]Vehkaoja A, Verho J, Comert A, Aydogan B, Perhonen M, Lekkala J, Halttunen J: System for ECG and heart rate monitoring during group training. In Engineering in Medicine and Biology Society. EMBS: 30th Annual International Conference of the IEEE; 2008:4832-4835.
- [8]Catrysse M, Puers R, Hertleer C, Van Langenhove L, van Egmond H, Matthys D: Towards the integration of textile sensors in a wireless monitoring suit. Sensor Actuat A-Phys 2004, 114:302-311.
- [9]Muhlsteff J, Such O, Schmidt R, Perkuhn M, Reiter H, Lauter J, Thijs J, Musch G, Harris M: Wearable approach for continuous ECG - and activity patient-monitoring. IEEE 2004, 1:2184-2187.
- [10]Hoffmann K-P, Ruff R: Flexible dry surface-electrodes for ECG long-term monitoring. Conf Proc IEEE Eng Med Biol Soc 2007, 2007:5740-5743.
- [11]Vehkaoja A, Verho J, Cömert A, Honkala M, Lekkala J: Wearable System for EKG Monitoring - Evaluation of Night-Time Performance. In Wireless mobile communication and healthcare. 83rd edition. Edited by Nikita KS, Lin JC, Fotiadis DI, Arredondo Waldmeyer M-T B. Berlin Heidelberg: Heidelberg: Springer; 2012:119-126.
- [12]Puurtinen MM, Komulainen SM, Kauppinen PK, Malmivuo JAV, Hyttinen JAK: Measurement of noise and impedance of dry and wet textile electrodes, and textile electrodes with hydrogel. Conf Proc IEEE Eng Med Biol Soc 2006, 1:6012-6015.
- [13]Scilingo EP, Gemignani A, Paradiso R, Taccini N, Ghelarducci B, DeRossi D: Performance evaluation of sensing fabrics for monitoring physiological and biomechanical variables. IEEE Trans Inf Technol Biomed 2005, 9:345-352.
- [14]Such O, Muehlsteff J: The Challenge of Motion Artifact Suppression in Wearable Monitoring Solutions. Medical Devices and Biosensors, 2006. 3rd IEEE/EMBS International Summer School; 2006:49-52. http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumber=4201263 webcite
- [15]Dozio R, Baba A, Assambo C, Burke MJ: Time based measurement of the impedance of the skin-electrode interface for dry electrode ECG recording. Conf Proc IEEE Eng Med Biol Soc 2007, 2007:5001-5004.
- [16]Romero I, Berset T, Buxi D, Brown L, Penders J, Kim S, Van Helleputte N, Kim H, Van Hoof C, Yazicioglu F: Motion artifact reduction in ambulatory ECG monitoring: an integrated system approach. In Proceedings of the 2nd Conference on Wireless Health. New York, NY, USA: ACM; 2011:11:1-11:8.
- [17]Kirst M, Lamparth S, Fuhrhop S, Ottenbacher J, Kunze C: Transferring an Artifact Handling System into a Mobile Long-term ECG Device with Dry Electrodes. In World Congress on Medical Physics and Biomedical Engineering, September 7–12, 2009, Munich, Germany. Edited by Dössel O, Schlegel WC. Berlin Heidelberg: Magjarevic R Springer; 2010:942-945.
- [18]Ottenbacher J, Kirst M, Jatoba L, Huflejt M, Grossmann U, Stork W: Reliable motion artifact detection for ECG monitoring systems with dry electrodes. IEEE; 2008:1695-1698. [30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2008. EMBS 200]
- [19]Kim S, Leonhardt S, Zimmermann N, Kranen P, Kensche D, Muller E, Quix C: Influence of contact pressure and moisture on the signal quality of a newly developed textile ECG sensor shirt. Proceedings of the 5th International Workshop on Wearable and Implantable Body Sensor Networks; 2008:256-259. http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumber=4575068 webcite
- [20]Wiese SR, Anheier P, Connemara RD, Mollner AT, Neils TF, Kahn JA, Webster JG: Electrocardiographic motion artifact versus electrode impedance. Biomedical Engineering, IEEE Transactions on 2005, 52:136-139.
- [21]Kirst M, Glauner B, Ottenbacher J: Using DWT for ECG motion artifact reduction with noise-correlating signals. Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE; 2011:4804-4807. http://ieeexplore.ieee.org/xpl/articleDetails.jsp?arnumber=6091190 webcite
- [22]Kang T-H, Merritt CR, Grant E, Pourdeyhimi B, Nagle HT: Nonwoven fabric active electrodes for biopotential measurement during normal daily activity. Biomedical Engineering, IEEE Transactions on 2008, 55:188-195.
- [23]Jarno M, Tanskanen JJV: A. Tanskanen JJV: Independent Component Analysis in ECG Signal Processing; 2012.
- [24]Milanesi M, Martini N, Vanello N, Positano V, Santarelli MF, Landini L: Independent component analysis applied to the removal of motion artifacts from electrocardiographic signals. Med Biol Eng Comput 2008, 46:251-261.
- [25]Hamilton PS, Curley MG, Aimi RM, Sae-Hau C: Comparison of methods for adaptive removal of motion artifact. Computers in Cardiology 2000, 2000:383-386.
- [26]Liu Y, Pecht MG: Reduction of skin stretch induced motion artifacts in electrocardiogram monitoring using adaptive filtering. Conf Proc IEEE Eng Med Biol Soc 2006, 1:6045-6048.
- [27]de Talhouet H, Webster JG: The origin of skin-stretch-caused motion artifacts under electrodes. Physiol Meas 1996, 17:81-93.
- [28]Medrano G, Ubl A, Zimmermann N, Gries T, Leonhardt S: Skin Electrode Impedance of Textile Electrodes for Bioimpedance Spectroscopy. In 13th International Conference on Electrical Bioimpedance and the 8th Conference on Electrical Impedance Tomography. Edited by Scharfetter H, Merwa R, Magjarevic R. Berlin Heidelberg: Springer; 2007:260-263. http://eu.wiley.com/WileyCDA/WileyTitle/productCd-EHEP000554.html webcite
- [29]Neuman MR: Biopotential Electrodes. In Medical Instrumentation: Application and Design. 3rd edition. Edited by Webster JG. John Wiley & Sons Inc; 1997:183-232.
- [30]Beckmann L, Neuhaus C, Medrano G, Jungbecker N, Walter M, Gries T, Leonhardt S: Characterization of textile electrodes and conductors using standardized measurement setups. Physiol Meas 2010, 31:233-247.
- [31]Pan J, Tompkins WJ: A real-time QRS detection algorithm. Biomedical Engineering, IEEE Transactions on 1985, BME-32:230-236.
- [32]Odman S: Changes in skin potentials induced by skin compression. Med Biol Eng Comput 1989, 27:390-393.
- [33]Vos WK, Bergveld P, Marani E: Low frequency changes in skin surface potentials by skin compression: experimental results and theories. Arch Physiol Biochem 2003, 111:369-376.
- [34]Spach MS, Barr RC, Havstad JW, Long EC: Skin-electrode impedance and its effect on recording cardiac potentials. Circulation 1966, 34:649-656.
- [35]Pengjun X, Xiaoming T, Shanyuan W: Measurement of wearable electrode and skin mechanical interaction using displacement and pressure sensors. Biomedical Engineering and Informatics (BMEI), 2011 4th International Conference on 2011, 2:1131-1134.