BioMedical Engineering OnLine
Impedance spectroscopy of changes in skin-electrode impedance induced by motion
Alper Cömert1  Jari Hyttinen1 
[1] Tampere University of Technology, Department of Electronics and Communications Engineering, BioMediTech, Tampere, Finland
关键词: EMG;    ECG;    Textile electrodes;    Surface electrodes;    Motion artifact;    Skin-electrode impedance;   
Others  :  1084237
DOI  :  10.1186/1475-925X-13-149
 received in 2014-06-19, accepted in 2014-11-03,  发布年份 2014
PDF
【 摘 要 】

Background

The motion artifact is an ever-present challenge in the mobile monitoring of surface potentials. Skin-electrode impedance is investigated as an input parameter to detect the motion artifact and to reduce it using various methods. However, the impact of the used impedance measurement frequency on the relationship between measured impedance and the motion artifact and the relationship between the impedance and the motion is not well understood.

Methods

In this paper, for the first time, we present the simultaneous measurement of impedance at 8 current frequencies during the application of controlled motion to the electrode at monitored electrode mounting force. Three interwoven frequency groupings are used to obtain a spectrum of 24 frequencies between 25 Hz and 1 MHz for ten volunteers. Consequently, the surface potential and one channel of ECG are measured from the electrode subject to controlled motion. The signals are then analyzed in time and frequency domain.

Results

The results show that the different frequencies of impedance measurements do not reflect the motion in the same manner. The best correlation between impedance and the applied motion was seen at impedance current frequencies above 17 kHz. For resistance this relationship existed for frequencies above 11 kHz, Reactance did not show good time domain correlation, but had good frequency domain correlation at frequencies higher than 42 kHz. Overall, we found that the impedance signal correlated well with the applied motion; however impedance had lower correlation to actual motion artifact signal.

Conclusion

Based on our results, we can conclude that the current frequency used for the impedance measurement has a great effect on the relationship of the measurement to the applied motion and its relationship with the resulting motion artifact. Therefore, when flat textile contact biopotential electrodes are used, frequencies higher than 17 kHz are best suited for impedance measurements intended for the estimation of electrode motion or motion artifact. For resistance, the best frequencies to use are higher than 11 kHz.

【 授权许可】

   
2014 Cömert and Hyttinen; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150113155927643.pdf 1751KB PDF download
Figure 9. 147KB Image download
Figure 8. 153KB Image download
Figure 7. 125KB Image download
Figure 6. 76KB Image download
Figure 5. 120KB Image download
Figure 4. 134KB Image download
Figure 3. 166KB Image download
Figure 2. 263KB Image download
Figure 1. 127KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

Figure 4.

Figure 5.

Figure 6.

Figure 7.

Figure 8.

Figure 9.

【 参考文献 】
  • [1]Hamilton PS, Curley MG, Aimi RM, Sae-Hau C: Comparison of methods for adaptive removal of motion artifact. In Comput Cardiol 2000. New York City, USA: IEEE; 2000:383-386.
  • [2]Liu Y: Reduction of skin stretch induced motion artifacts in electrocardiogram monitoring using adaptive filtering. PhD thesis. College Park, Maryland, USA: University of Maryland, Department of Mechanical Engineering; 2007.
  • [3]Pengjun X, Xiaoming T, Shanyuan W: Measurement of wearable electrode and skin mechanical interaction using displacement and pressure sensors. 4th Int Conf Biomed Eng Inform BMEI. Volume 2 2011, 1131-1134.
  • [4]Tong DA, Bartels KA, Honeyager KS: Adaptive reduction of motion artifact in the electrocardiogram. In 24th Annu Conf Annu Fall Meet Biomed Eng Soc EMBSBMES Conf 2002 Proc Second Jt. Volume 2. New York City, USA: IEEE; 2002:1403-1404. http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumber=1106451 webcite
  • [5]Raya MA, Sison LG: Adaptive noise cancelling of motion artifact in stress ECG signals using accelerometer. In Eng Med Biol 2002 24th Annu Conf Annu Fall Meet Biomed Eng Soc EMBSBMES Conf 2002 Proc Second Jt. Volume 2. New York City, USA: IEEE; 2002:1756-1757. vol.2. http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumber=1106637 webcite
  • [6]Romero I, Berset T, Buxi D, Brown L, Penders J, Kim S, Van Helleputte N, Kim H, Van Hoof C, Yazicioglu F: Motion Artifact Reduction in Ambulatory ECG Monitoring: An Integrated System Approach. In Proc 2Nd Conf Wirel Health. New York, NY, USA: ACM; 2011. 11:1–11:8. [WH ’11]
  • [7]Lee W-C, Yang Y-SO, Ke T-C, Wei C-S, Lee H-C: Adaptive reduction of motion artifact in a portable ECG system. In 2010 IEEE Sens. New York City, USA: IEEE; 2010:704-707. http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumber=5690058 webcite
  • [8]Hamilton PS, Curley MG: Adaptive Removal of Motion Artifact. In Proc 19th Annu Int Conf IEEE Eng Med Biol Soc 1997. Volume 1. New York City, USA: IEEE; 1997:297-299. http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumber=754531 webcite
  • [9]Comert A, Honkala M, Hyttinen J: Effect of pressure and padding on motion artifact of textile electrodes. Biomed Eng OnLine 2013, 12:26. BioMed Central Full Text
  • [10]Ottenbacher J, Kirst M, Jatobá L, Großmann U, Stork W: An approach to reliable motion artifact detection for mobile long-term ECG monitoring systems using dry electrodes. In IV Lat Am Congr Biomed Eng 2007 Bioeng Solut Lat Am Health. Edited by Müller-Karger C, Wong S, Cruz AL. Berlin Heidelberg: Springer; 2008:440-443. [IFMBE Proceedings, vol. 18]
  • [11]Buxi D, Kim S, van Helleputte N, Altini M, Wijsman J, Yazicioglu RF, Penders J, Van Hoof C: Correlation between electrode-tissue impedance and motion artifact in biopotential recordings. IEEE Sens J 2012, 12:3373-3383.
  • [12]Ödman S, Öberg PÅ: Movement-induced potentials in surface electrodes. Med Biol Eng Comput 1982, 20:159-166.
  • [13]Ödman S: Potential and impedance variations following skin deformation. Med Biol Eng Comput 1981, 19:271-278.
  • [14]de Talhouet H, Webster JG: The origin of skin-stretch-caused motion artifacts under electrodes. Physiol Meas 1996, 17:81-93.
  • [15]Beckmann L, Neuhaus C, Medrano G, Jungbecker N, Walter M, Gries T, Leonhardt S: Characterization of textile electrodes and conductors using standardized measurement setups. Physiol Meas 2010, 31:233.
  • [16]Albulbul A, Chan ADC: Electrode-skin impedance changes due to an externally applied force. In 2012 IEEE Int Symp Med Meas Appl Proc MeMeA. New York City, USA: IEEE; 2012:1-4. http://ieeexplore.ieee.org/xpl/articleDetails.jsp?arnumber=6226628 webcite
  • [17]Vos WK, Bergveld P, Marani E: Low frequency changes in skin surface potentials by skin compression: experimental results and theories. Arch Physiol Biochem 2003, 111:369-376.
  • [18]Rosell J, Colominas J, Riu P, Pallas-Areny R, Webster JG: Skin impedance from 1 Hz to 1 MHz. IEEE Trans Biomed Eng 1988, 35:649-651.
  • [19]Lin C-T, Liao L-D, Liu Y-H, Wang I-J, Lin B-S, Chang J-Y: Novel Dry polymer foam electrodes for long-term EEG measurement. IEEE Trans Biomed Eng 2011, 58:1200-1207.
  • [20]Gruetzmann A, Hansen S, Müller J: Novel dry electrodes for ECG monitoring. Physiol Meas 2007, 28:1375.
  • [21]Hoffmann K-P, Ruff R: Flexible dry surface-electrodes for ECG long-term monitoring. Conf Proc Annu Int Conf IEEE Eng Med Biol Soc IEEE Eng Med Biol Soc Conf 2007, 2007:5740-5743.
  • [22]Lymed Product Catalog. Tampere, Finland: Lymed OY; [http://www.lymed.fi/wp-content/uploads/2014/06/Kuvasto_web_2.1_ENG.pdf webcite]
  • [23]Meziane N, Webster JG, Attari M, Nimunkar AJ: Dry electrodes for electrocardiography. Physiol Meas 2013, 34:R47.
  文献评价指标  
  下载次数:136次 浏览次数:27次