期刊论文详细信息
BioMedical Engineering OnLine
Selected parameters of the corneal deformation in the Corvis tonometer
Robert Koprowski3  Anita Lyssek-Boron2  Anna Nowinska2  Edward Wylegala2  Henryk Kasprzak1  Zygmunt Wrobel3 
[1] Institute of Physics, Wroclaw University of Technology, Wybrzeze Wyspianskiego 27, Wroclaw 50-370, Poland
[2] Ophthalmology Clinic, Medical University of Silesia, District Railway Hospital in Katowice, Katowice, Poland
[3] Department of Biomedical Computer Systems, University of Silesia, Faculty of Computer Science and Materials Science, Institute of Computer Science, ul. Będzińska 39, Sosnowiec 41-200, Poland
关键词: Segmentation;    Measurement Automation;    Image processing;    Corvis tonometer;    Eye biomechanics;   
Others  :  793494
DOI  :  10.1186/1475-925X-13-55
 received in 2014-03-19, accepted in 2014-04-29,  发布年份 2014
PDF
【 摘 要 】

Introduction

Contemporary ophthalmology knows many methods of measuring intraocular pressure, namely the methods of non-contact and impression applanation tonometry. In non-contact applanation tonometers, e.g. the Corvis, the corneal flattening is caused by an air puff. Image registration of the corneal deflection performed by a tonometer enables to determine other interesting biomechanical parameters of the eye, which are not available in the tonometer. The measurement of new selected parameters is presented in this paper.

Material and method

Images with an M × N × I resolution of 200 × 576 × 140 pixels were acquired from the Corvis device in the source recording format *.cst. A total of 13'400 2D images of patients examined routinely in the Clinical Department of Ophthalmology, in District Railway Hospital in Katowice, Poland, were analysed in accordance with the Declaration of Helsinki. A new method has been proposed for the analysis of corneal deflection images in the Corvis tonometer with the use of the Canny edge detection method, mathematical morphology methods and context-free operations.

Results

The resulting image analysis tool allows determination of the response of the cornea and the entire eyeball to an air puff. The paper presents the method that enables the measurement of the amplitude of curvature changes in the frequency range from 150 to 500 Hz and automatic designation of the eyeball movement direction. The analysis of these data resulted in 3 new features of dynamics of the eye reaction to an air puff. Classification of these features enabled to propose 4 classes of deformation. The proposed algorithm allows to obtain reproducible results fully automatically at a time of 5 s per patient using the Core i5 CPU M460 @ 2.5GHz 4GB of RAM.

Conclusions

The paper presents the possibility of using a profiled algorithm of image analysis, proposed by the authors, to measure additional cornea deformation parameters. The new tool enables automatic measurement of the additional new parameters when using the Corvis tonometer. A detailed clinical examination based on this method will be presented in subsequent papers.

【 授权许可】

   
2014 Koprowski et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20140705052259291.pdf 1600KB PDF download
Figure 13. 60KB Image download
Figure 12. 31KB Image download
Figure 11. 108KB Image download
Figure 10. 67KB Image download
Figure 9. 56KB Image download
Figure 8. 131KB Image download
Figure 7. 81KB Image download
Figure 6. 67KB Image download
Figure 5. 65KB Image download
Figure 4. 65KB Image download
Figure 3. 65KB Image download
Figure 2. 73KB Image download
Figure 1. 35KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

Figure 4.

Figure 5.

Figure 6.

Figure 7.

Figure 8.

Figure 9.

Figure 10.

Figure 11.

Figure 12.

Figure 13.

【 参考文献 】
  • [1]Luce DA: Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg 2005, 31(1):156-162.
  • [2]Kobayashi AS, Staberg LG, Schlegel WA: Viscoelastic properties of human cornea. Exp Mech 1973, 12:497-503.
  • [3]Malik NS, Moss SJ, Ahmed N, Furth AJ, Wall RS, Meek KM: Ageing of the human corneal stroma: structural and biochemical changes. Biochim Biophys Acta 1992, 1138(3):222-228.
  • [4]Sherrard ES, Novakovic P, Speedwell L: Age-related changes of the corneal endothelium and stroma as seen in vivo by specular microscopy. Eye (Lond) 1987, 1(Pt2):197-203.
  • [5]Kotecha A, Elsheikh A, Roberts CR, Zhu HG, Garway-Heath DF: Corneal thicknessand age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci 2006, 47(12):5337-5347.
  • [6]Valbon BF, Ambrósio-Jr R, Fontes BM, Alves MR: Effects of age on corneal deformation by non-contact tonometry integrated with an ultra-high-speed (UHS) Scheimpflug camera. Arq Bras Oftalmol 2013, 76(4):229-232.
  • [7]Brown KE, Congdon NG: Corneal structure and biomechanics: impact on the diagnosis and management of glaucoma. Curr Opin Ophthalmol 2006, 17(4):338-343.
  • [8]Dupps WJ, Wilson SE: Biomechanics and wound healing in the cornea. Exp Eye Res 2006, 83(4):709-720.
  • [9]Elsheikh A, Alhasso D, Rama P: Assessment of the epithelium’s contribution to corneal biomechanics. Exp Eye Res 2008, 86(2):445-451.
  • [10]Meek KM, Leonard DW: Ultrastructure of the corneal stroma: a comparative study. Biophys J 1993, 64(1):273-280.
  • [11]Fontes BM, Ambrosio R Jr, Alonso RS, Jardim D, Velarde GC, Nose W: Corneal biomechanical metrics in eyes with refraction of −19.00 to +9.00 D in healthy brazilian patients. J Refract Surg 2008, 24(9):941-945.
  • [12]Kirwan C, O’Keefe M, Lanigan B: Corneal hysteresis and intraocular pressure measurement in children using the Reichert ocular response analyzer. Am J Ophthalmol 2006, 142(6):990-992.
  • [13]Shah S, Laiquzzaman M, Bhojwani R, Mantry S, Cunliffe I: Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes. Invest Ophthalmol Vis Sci 2007, 48(7):3026-3031.
  • [14]Touboul D, Roberts C, Kerautret J, Garra C, Maurice-Tison S, Saubusse E, Colin J: Correlation between corneal hysteresis intraocular pressure, and corneal central pachymetry. J Cataract Refract Surg 2008, 34(4):616-622.
  • [15]Congdon NG, Broman AT, Bandeen-Roche K, Grover D, Quigley HA: Central corneal thickness and corneal hysteresis associated with glaucoma damage. Am J Ophthalmol 2006, 141(5):868-875.
  • [16]Gatinel D, Chaabouni S, Adam PA, Munck J, Puech M, Hoang-Xuan T: Corneal hysteresis, resistance factor, topography, and pachymetry fter corneal lamellar flap. J Refract Surg 2007, 23(1):76-84.
  • [17]Shah S, Laiquzzaman M, Cunliffe I, Mantry S: The use of the Reichert ocular response analyser to establish the relationship between ocular hysteresis, corneal resistance factor and central corneal thickness in normal eyes. Cont Lens Anterior Eye 2006, 29(5):257-262.
  • [18]Ortiz D, Pinero D, Shabayek MH, Arnalich-Montiel F, Alió JL: Corneal biomechanical properties in normal, post-laser in situ keratomileusis, and keratoconic eyes. J Cataract Refract Surg 2007, 33(8):1371-1375.
  • [19]Pepose JS, Feigenbaum SK, Qazi MA, Sanderson JP, Roberts CJ: Changes in corneal biomechanics and intraocular pressure following LASIK using static, dynamic, and noncontact tonometry. Am J Ophthalmol 2007, 143(1):39-47.
  • [20]Elsheikh A, Anderson K: Comparative study of corneal strip extensometry and inflation tests. J R Soc Interface 2005, 2(3):177-185.
  • [21]Reznicek L, Muth D, Kampik A, Neubauer AS, Hirneiss C: Evaluation of a novel Scheimpflug–based non–contact tonometer in healthy subjects and patents with ocular hypertension and glaucoma. Br J Ophthalmol 2013, 97(11):1410-1414.
  • [22]Hon Y, Lam AKC: Corneal deformation measurement using scheimpflug noncontact tonometer. Optometry & Vistion Science 2013, 90(1):e1-e8.
  • [23]Leung CKS, Ye C, Weinreb RN: An ultra high-speed Scheimpflug camera for evaluation of corneal deformation response and its impact on IOP measurement. Investigative Ophtalmology & Visual Science 2013, 54(4):2885-2892.
  • [24]Kling S, Marcos S: Contributing factors to corneal deformation in air puff measurements. Investigative Ophtalmology & Visual Science 2013, 54(7):5078-5085.
  • [25]Otsu N: A threshold selection method from gray-level histograms. IEEE Trans Sys, Man, Cyber 1979, 9(1):62-66.
  • [26]Koprowski R, Wrobel Z: Identification of layers in a tomographic image of an eye based on the canny edge detection. Inf Technol Biomed Adv Intell Soft Comput 2008, 47:232-239.
  • [27]Koprowski R, Wróbel Z: Layers recognition in tomographic eye image based on random contour analysis. Computer recognition systems 3. Adv Intell Soft Comput 2009, 57:471-478.
  • [28]Korzynska A, Iwanowski M: Multistage morphological segmentation of bright-field and fluorescent microscopy images. Opt-Electron Rev 2012, 20(2):87-99.
  • [29]Wilczyński S, Ptaszkiewicz M, Pierzchała E, Pilawa B, Swakoń J, Olko P: Electron paramagnetic resonance studies of gamma irradiated azithromycin. Current Topics in Biophysics 2008, 31:1-4.