会议论文详细信息
10th Asia Plasma and Fusion Association Conference | |
Preliminary Results from Electron Cyclotron Measurements at SST-1 | |
Siju, Varsha^1 ; Shukla, Praveena^1 ; Raval, Jayesh^1 ; Shankara, Joisa Y.^1 ; Pathak, S.K.^1 | |
Institute for Plasma Research, Bhat, Gandhinagar | |
382 428, India^1 | |
关键词: Electron cyclotron emission; Electron cyclotrons; Heterodyne radiometer; Measurements of; Noise temperature; Radiation temperature; Second harmonics; Spatial resolution; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/823/1/012057/pdf DOI : 10.1088/1742-6596/823/1/012057 |
|
来源: IOP | |
【 摘 要 】
An 8-channel heterodyne radiometer system is developed and installed for the measurements of second harmonic electron cyclotron emission (ECE) at magnetic field of 1.5 T at SST-1. This system covers a spectral range of 75.4 to 84.5 GHz at a spatial resolution of less than 1 cm, sensitivity of 9.51 x 106V/W. The calculated noise temperature of the system is 1.66 eV. The system is calibrated using Hot/cold technique. This paper presents the preliminary observations of the heterodyne radiometer system at SST-1 and the measured radiation temperature.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
Preliminary Results from Electron Cyclotron Measurements at SST-1 | 1415KB | download |